Guidelines and manufacturer's declaration – electromagnetic immunity
The blood pressure measuring device Mobil-O-Graph
The customer or user of the blood pressure measuring device Mobil-O-Graph
such an environment.
Immunity test
Test level
Electrostatic
+- 8 kV contact discharge
discharge (ESD)
+- 15 kV air discharge
according to IEC
61000-4-2
Fast transient
± 1 kV
electrical
100 kHz repetition rate
disturbance/bursts
according to IEC
61000-4-4
Surge voltages
(surges) according to
IEC 61000-4-5
68
is intended for use in the environment specified below.
®
Compliance level
+- 8 kV contact discharge
+- 15 kV air discharge
± 1 kV
100 kHz repetition rate
Not applicable
should ensure that it is used in
®
Electromagnetic environment –
guidelines
Floors should consist of wood
or cement or be covered with
ceramic tiles. If the floor
consists of synthetic materials,
relative humidity must be at
least 30 %.
The Mobil-O-Graph
does not
®
have an AC power supply