Specifying the desired measurement log

You have the option, using the HMS CS, to adjust the following settings for the measurement log:
Beginning of 4 different daily intervals
Number of measurements per hour in the 4 daily intervals
Enable/disable the acoustic signals for the daily intervals
Enable/disable the optional PWA
Choose between 24-hour ABPM and in-clinic monitoring
Turn the display on/off.
As soon as you have conducted a measurement, the log can only be changed once you have completely deleted
all data.
You can set the log manually by pressing and holding the day/night button while simultaneously pressing the event
button. Use the start/stop button to change the log and confirm with the event button.
Note:
To use practice monitoring a Bluetooth
supplier or directly from IEM GmbH.
Setting the logs via software
To set the protocols using the software please refer to the HMS CS instructions for use for the
respective patient data management software.
interface is required; you can obtain one of these via your specialist
®
Product description
EN
29

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Mobil-O-Graph and is the answer not in the manual?

Questions and answers

Audrey
March 20, 2025

What does Error 10 mean?

1 comments:
Mr. Anderson
March 20, 2025

Error 10 on the IEM Mobil-O-Graph indicates a serious malfunction due to pressure buildup outside the measurement process, meaning the pump was activated improperly. This error suggests a critical program code failure. The device should be sent immediately for inspection and repair to the dealer or directly to IEM GmbH.

This answer is automatically generated

This manual is also suitable for:

Mobil-o-graph pwaM26150200010100013070x02001

Table of Contents