Download Print this page

Permobil ROHO Dry FLOATATION QUADTRO SELECT E2624 Adjustment Manual page 3

Cushion

Advertisement

ROHO
Dry FLOATATION
®
inflation and adjustment guide
The QUADTRO SELECT has four compartments of air cells. When the ISOFLO
The inflation level of each compartment can be separately adjusted. This allows you to correct or accommodate side-to-side and/or anterior-posterior asymmetries of the pelvis and upper legs. NOTE: Before adjusting the cushion
for an asymmetry, you MUST adjust the air inflation level to the proper cushion inflation setting as outlined below.
QUADTRO SELECT setup
Inflate and adjust
the ROHO QUADTRO SELECT
cushion for a new individual.
1. Place the Quadtro Select cushion in the wheelchair
with the air cells facing up and the non-skid surface
of the cover facing down.
2. Before inflating, you must open the ISOFLO
Memory Control so that all of the air cells will
inflate. Push the ISOFLO green knob toward the
unlocked position.
3. Slide the rubber nozzle of the hand pump over the
inflation valve. Pinch the nozzle, turn to open the
inflation valve, and overinflate the air cells.
4.
Remove the hand pump.
Hand pump
permobil.com
cushion
®
Memory Control is open, the air flows freely between the compartments. When the ISOFLO is closed, the air is "locked" within each compartment.
®
QUADTRO SELECT adjustment
Check
the inflation level daily by performing a
hand check.
After overinflating the cushion, perform a hand check to
adjust to the proper cushion inflation setting:
1. The individual should sit in the chair in their usual
position. The cushion should be centered beneath
the individual.
2. Perform a hand check:
a. Slide your hand between
the cushion and the
individual. Lift the leg
slightly and feel for the
lowest bony prominence,
and then lower the leg to a
sitting position.
b. Open the valve to remove
Open ISOFLO -
Push green knob in
air, keeping your hand
beneath the individual's
Open
lowest bony prominence.
c. Release air until you
can slightly move your
Close
fingertips. Close the valve.
Hand check results
Re-adjust
(typically every 3-4 months).
• If you can slightly move your fingertips: the
cushion is at the Proper Cushion Inflation
Setting (½" - 1" air between the lowest bony
prominence and the seat surface) and no
further adjustment is needed.
• If you can easily move your fingertips: there
is too much air between the lowest bony part
prominence and the seat surface. Open the
inflation valve and slowly release air until you
can barely move your fingertips, then close
the valve.
• If you cannot move your fingertips at all:
there is too little/no air between the lowest
bony prominence and the seat surface. Open
the inflation valve and slowly add air until
you can slightly move your fingertips, then
close the valve.
See reverse side for instructions on how
to adjust QUADTRO SELECT for postural
asymmetries.
½"-1"
Proper cushion inflation setting
½" – 1" (1.5 cm – 2.5 cm)
the inflation level only as needed
MKTG0054-00

Advertisement

loading
Need help?

Need help?

Do you have a question about the ROHO Dry FLOATATION QUADTRO SELECT E2624 and is the answer not in the manual?

This manual is also suitable for:

Roho dry floatation quadtro select e2625