Page 1
Calibration Set CS 2172 X Instruction Manual Version 1.2.4; Code. No. 20 752 149...
Page 2
Mark on your equipment certifies that it meets requirements of all subjected EU regulations. Hereby, Metrel d.d. declares that the CS 2172 X is in compliance with subjected EU directive. The full text of the EU declaration of conformity is available at the following internet address https://www.metrel.si/DoC.
MultiservicerXA MI 3321, CE Multitester MI 2094, MI 3360 OmegaPAT XA, MI 3325 MultiServicerXD, MI 3394 CE MultitesterXA and similar appliance safety testers. For complete adjustment of these Metrel devices, CS 2171 and some other equipment are needed. It contains parts (resistors) for adjustment of various test functions like: Continuity resistance;...
CS 2172 X Calibration set Calibration set description 2 Calibration set description The Calibration set CS 2172 X is housed in a plastic case with handle. Front plate has a clear layout for easy access to a particular test group. Figure 2.1: Bottom panel layout of the calibrator CS 2172 X...
Page 6
CS 2172 X Calibration set Calibration set description Figure 2.2: Top panel layout of the calibrator CS 2172 X...
Page 7
Calibration set CS 2172 X consists of: Calibration unit CS 2172 X; 2 cascade banana test leads; 2 non-cascade banana test leads; 2 HV test probes METREL; 1 Schuco test cable 1 jumper; Figure 2.3: Calibration set CS 2172 X...
CS 2172 X Calibration set Warnings 3 Warnings In order to reach the highest level of operator’s safety while carrying out various tests when using the Calibration set CS 2172 X, as well as to ensure that the test equipment remains undamaged, it is necessary to consider the following warnings: Read the Instruction manual with special care for safe operation.
CS 2172 X Calibration set Operation with Calibration set 4 Operation with Calibration set Calibration set CS 2172 X contains set of 4 different function groups of resistors in bottom panel and 6 different function groups of resistors in top panel. Operator has to be familiar with the instrument adjustment procedure.
CS 2172 X Calibration set Maintenance 5 Maintenance 5.1 Inspection To maintain the operator’s safety and to ensure the reliability of the Calibration set it is advisable to inspect the set on a regular basis. Check that the calibrator CS 2172 X and its accessories are not damaged.
CS 2172 X Calibration set Technical specifications 6 Technical specifications Duty ratio 50% / 1 min max., except when specified different. 6.1 Bottom resistor groups Continuity & earth resistors group Rated resistors Tolerance (Reference Tolerance (Max. Max power [W] Value [] load <...
CS 2172 X Calibration set Technical specifications 6.2 Top resistor groups Continuity & earth resistors group Rated resistor Tolerance (Reference Tolerance (Max. Max power [W] Value [] load < 10 sec) load < 60 sec) ±(0.5 % + 2 mΩ) ±(1 % + 2 mΩ) 1.44 150 (10 A, 15V)
Page 13
CS 2172 X Calibration set Technical specifications Clamp resistors group Rated resistor Tolerance (Reference Tolerance (Max. Max power [W] Value [] load < 10 sec) load < 60 sec) ±(0.5 %) ±(1 %) 10 k Power resistors group Rated resistor Tolerance (Reference Tolerance (Max.
Page 14
CS 2172 X Calibration set Technical specifications General specifications Measuring category for Leakage and Power Resistor group ..........CAT II 300 V Protection classification ......... Class II Pollution degree ..........2 Dimensions (w × h × d) ........36 x 16 x 33 cm Mass (without accessories) ......
Need help?
Do you have a question about the CS 2172 Series and is the answer not in the manual?
Questions and answers