Eurocheck CS 2099 User manual Version 1.2, Code no. 20 750 127...
Page 2
Mark on your equipment certifies that this equipment meets the requirements of the EU (European Union) concerning safety and electromagnetic compatibility regulations. No part of this publication may be reproduced or utilized in any form or by any means without permission in writing from METREL.
1 Introduction to CS 2099 – Eurocheck CS 2099 – Eurocheck is a professional, multifunctional field calibrator intended for use with Metrel installation test instruments. In addition to its durable body and simple operation it boasts the following features: Insulation calibration with test voltages up to 1000 V, LowΩ...
CS 2099 Eurocheck Safety and operational considerations 2 Safety and operational considerations 2.1 Warnings Read and understand all warnings and cautions before using this product. This will ensure your safety, as well as keep the test equipment undamaged. symbol on the instrument means »Read the Instruction manual with special care«.
CS 2099 Eurocheck Eurocheck description and maintenance 3 Eurocheck description and maintenance 3.1 Front panel Note: Output connector Figure 3.1: Front panel type is country dependent. Legend: 1 ...Insulation calibration outputs. 2 ...LowΩ and continuity calibration outputs. Caution – do not use test currents higher than 200 mA. 3 ...Power supply cord connection polarity indication.
CS 2099 Eurocheck Eurocheck description and maintenance 3.2 Side panel Note: Output connector type is country dependent. Figure 3.2: Side panel Legend: 1 ...Power supply cord. 2 ...Fuse F1. Note: Eurocheck powers on automatically 2 seconds after it is connected to the mains supply. It checks for correct mains supply plug orientation during power up.
CS 2099 Eurocheck Eurocheck description and maintenance 3.3 Replacing fuse A fuse is located under side cover of the Eurocheck unit: T 4 A / 230 V, 32 mm × 6.3 mm General input protection fuse. Warnings: Disconnect any connected instrument/accessory and power off the calibrator before opening fuse compartment cover, hazardous voltage inside! Replace blown fuse with original type only, otherwise the calibrator may be damaged and/or operator’s safety impaired!
CS 2099 Eurocheck Eurocheck operation 4 Eurocheck operation 4.1 Recommended calibration sequence and practice This chapter lists recommended sequence for calibration of installation test instruments. It is not obligatory to follow this sequence. Eurocheck offers additional calibration functions (all functions are explained in detail in the following chapters). Calibration using universal test cable Step Calibrated function...
Eurocheck operation 4.2 Sample test report form Caution: This sample test report form is suitable for Metrel installation test instruments like MI3100 – EurotestEASI. It could be easily modified to suit other types of instruments. It is not based on any regulation or recommendation.
CS 2099 Eurocheck Eurocheck operation 4.3 Selecting calibration function/sub-function The following measurements can be selected with the RCD/LOOP function selection button (see Figure 3.1). Appropriate LED lights when particular function/sub-function is selected: Fault loop and trip-lock impedance calibration (two impedance values available), RCD trip-out time calibration and instrument test current verification.
CS 2099 Eurocheck Eurocheck operation 4.5 Fault loop impedance calibration Fault loop impedance is measured between L and PE conductors. Fault loop calibration will not trip an RCD if it’s present in installation where Eurocheck is connected, regardless of selected IΔn current. Caution: When performing fault loop impedance calibration Eurocheck connects fault loop calibration output to mains line impedance (L and N, important when...
Page 13
CS 2099 Eurocheck Eurocheck operation Note: In some instruments the function is named “Fault loop resistance”; however the calibration procedure is the same. Figure 4.1: Performing fault loop impedance calibration using plug cable Figure 4.2: Performing fault loop impedance calibration using universal test cable...
CS 2099 Eurocheck Eurocheck operation 4.6 Trip-lock impedance calibration Fault loop impedance is measured between L and PE conductors. Trip-lock calibration will not trip an RCD if it’s present in installation where Eurocheck is connected, regardless of selected IΔn current. Caution: When performing fault loop impedance calibration Eurocheck connects fault loop calibration output to mains line impedance (L and N, important when...
CS 2099 Eurocheck Eurocheck operation Selected IΔn current will affect dissipation of measurement results. Lower IΔn currents (especially 10 mA) will cause larger variations in results. Consider accuracy of trip-lock function when evaluating calibration results (accuracy is defined in User manual of your instrument). 4.7 Line impedance calibration Line impedance is measured between L and N conductors.
CS 2099 Eurocheck Eurocheck operation 4.8 RCD trip-out time calibration Eurocheck simulates an RCD with IΔn of 30 mA and trip-out time of 50 ms. Trip-out time calibration will not trip an RCD if it’s present in installation where Eurocheck is connected, regardless of selected IΔn current.
It is essential that the test instrument measuring current is within certain limits. For example, standard EN 61557 defines low limit as IΔn – 0 %, and high limit as IΔn + 20 %. Metrel test instruments has even narrower specification, high limit is defined as IΔn + 10 %.
– 0 Ω (short-circuited test leads). Metrel installation test instruments usually offer 2 distinctive functions for measuring low resistances. LowΩ usually stands for measuring low resistances with 200 mA test current, often with automatic polarity reversal. Continuity denotes a function with lower measuring current, usually around 7 mA.
Page 19
Notes: Most Metrel installation test instruments use L and N measuring leads of universal test cable to perform low resistance measurements. However, there are exceptions to this rule like MI 3101 – EurotestAT, which uses N and PE leads.
– 0 V (short-circuited test leads, not open). Frequency is equal to mains and is usually very close to 50 Hz or 60 Hz. Most Metrel installation test instruments measure all three line voltages: U L-PE .
CS 2099 Eurocheck Eurocheck operation Figure 4.5: Performing voltage and frequency calibration using universal test cable 4.13 Continuity and insulation of measurement leads verification Continuity of measurement leads Continuity of universal test cable measurement leads could be verified by voltage calibration between all 3 installation wires (between L and N, L and PE, N and PE).
CS 2099 Eurocheck Eurocheck operation Insulation of measurement leads Insulation of measurement leads may deteriorate over time. This can cause measurement and/or safety related problems. If insulation material leaks current measured results will be wrong. Also in such case there is a risk of electric shock when the user touches deteriorated part of insulation.
CS 2099 Eurocheck Technical specifications 5 Technical specifications 5.1 General Power supply voltage ......240 V a.c. (207 V/255 V), 50 Hz/60 Hz Protection classification..... double insulation Pollution degree ........ 2 Protection degree......IP 40 Overvoltage category ......CAT II 300 V Power consumption......
Page 24
CS 2099 Eurocheck Technical specifications Voltage and frequency Reference value Accuracy ± (5 % + accuracy of mains 50.0 V @ 240 V a.c. 47.9 V @ 230 V a.c. voltage) Short-circuit current: <6 mA. Nominal output resistance: 25 kΩ. Specified accuracy is valid only if Eurocheck output is enabled.