Description
1 lOOC-Series
Oscilloscope
mainframe
that accommodates
amplifiers
Power Module
Leveled Sine Wave
Generators
Power Supply
DC Voltage Calibrator
(optional)
Digital Multimeter
(w/test leads)
Signal Standardizer
2-4
functional test. Procedure steps are based on the test equipment examples given,
but otherequipment with similar specifications may be substituted. Test results,
setup information, and related oorurectors and adapters may be altered if you use
different
equipment.
Table Z-2 - Test Equipment
Test Equipment
accommodates
amplifiers
Tektronix four-compart-
TEKTRONIX TM 504 Power Module
Leveled variable
Wave Generator with a TM SO&Series
Power Module
amplitude, 50 kHz or
6 MHz reference
TEKTRONIX SO 603 Leveled Sine
Leveled variable
Wave Generator with a TM B&Series
amplitude, 50 kHz or
Power Module
MHz
6
reference
Continuously variable from
Supply with a TM SOItSeries Power
adjustable from o-400
overcurrent
protection
output, O-4 v
Data Precision 8200
Amplitude: 250 mv
Rise rime: sf25 ps
Aberrations: < 1%
Fluke 8842A Digital Multimeter
Accuracy <O.Of 96
TEKTRONIX 067-0587-02
Tektronix Calibration
Fixture with interlace
Standardizer
connector modiiiecl for
1 loo&Series
use
lists recommenoed
test equipment for use In this
column of,Table 2-2 indicates,l with a check mark
of Recommended
Programmable 0scilloscope
11301
Programmable Oscilloscope
11301A Counter Timer Oscilloscope
Digitizing Oscilloscope
Tunnel diode
Signal
Checks and Adjusrmenrs
Test