I
Procedure to Check Amplliiet Bandwidth:
q
Step 1:
First Inltlallze the mainframe's settings, then perform the following
settings in the order listed:
CH 1 OisplayorWl .
Leveled sine wave generator
Frequency . . . . . . . _ . . . . . . . . . . . . , . + *, . . , . . . . . Ref (6 MHz)
Mainframe
Impedance . . . . . . . . . . . . . . . . . . . . . _ . . . . . 50 n
HORIZONTAL WE _ . . . . . . . , . . . . . . . , . , . . . . . , . I * , 1 , ,
Select the CURSORS to measure the peak-tc-peak amplitude of the
waveform.
Set the leveled sine wave generator for 1.2 V peak-m-peak.
q
Step 4:
Sat the leveled sine wava generator Frequency control to the
maximum bandwidth frequency specified for
combination. This limtt is displayed as HF Limit PERTICAL SIZE button).
q
Set the CH 1 display on/off button to off.
the coaxial cable to the CH 2 input and set its display on/off
to on.
q
Step 9:
Perform the Steps 1 to 8 for the remaining channels.
Functional Test Procedure
.
.
.
.
.
.
.
.
.
.
.
on
100 ns/div
2 - 2 7