Download Print this page

Tektronix 11A34 Service Reference Manual page 16

Four-channel amplifier
Hide thumbs Also See for 11A34:

Advertisement

q
'Step 7: Set the leveled sine wave generator fmquency to each Test
Frequency in mlumn (1) of Table Z-3, and record the Displayed Amplitude ii .,
Check Ainplifrar Bandwidth for CalculatinQ'the
may require more man one leveled alne wave generator to
Test Prequencies listed. When changing to another leveled sine
wave generator, repeat Steps 5 and 6.
Set the maInframe ON/STANDBY switch to STANDBY
q
Step 9:
compartment. Remove the amplifier from the Left plug-in compartment and
install it in
Center plug-in compartment.
me
Setup
to Examlne/AdJust
(AlR1027, AlR1029, AlR1041,
Procedure to Examine/Adjust Amplitter
Perform the following settings in the order listed:
Remove the left side cover from the amplifier.
n
Insert the amplifier into me mainframe Center plug-in compartment.
Connect me pulser to the CH 1 input connector
Connect a 50 D coaxial cable from the calibration generator High Amp1
l
connector to the pulser.
Calibration
generator
switch
.
Period
.
,
.
Pulse
Amplitude
Triggered
amplifier bandwidth.
AlR1043)
Step Response
.
.
.
.
,
.
.
.
,
.
.
.
.
.
.
,
.
Level
fully
Response:
High
.
0.1
. Max
clockwise

Advertisement