NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects, and people
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios,
amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the
electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the
measured field strength in the location in which the Pacifier Activated Lullaby (P.A.L) Model #PAL-HOS is used exceeds the
applicable RF compliance level above, the Model should be observed to verify normal operation. If abnormal performance is
observed, additional measures may be necessary, such as re-orienting or relocating the Pacifier Activated Lullaby (P.A.L) Model
# PAL-HOS. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m
Recommended separation distances between portable and mobile RF communications equipment and the
The Pacifier Activated Lullaby (P.A.L) Model PAL-HOS is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The user of the Pacifier Activated Lullaby (P.A.L) Model # PAL-HOS can help prevent electromagnetic
interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the
Pacifier Activated Lullaby (P.A.L) Model # PAL-HOS as recommended below, according to the maximum output power of the
communications equipment
Rated maximum output
power of transmitter
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be
determined using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people
© 2023 by NeoLight LLC.
All rights reserved.
Pacifier Activated Lullaby (
Separation distance according to frequency of transmitter m
d=1.2sqrt(P)
0.12
0.38
1.2
3.8
12
equation applicable to the
frequency of the transmitter
Recommended separation distance
d=1.2sqrt(P), 80 Mhz to 800 Mhz
d=2.3sqrt(P), 800 Mhz to 2.5 Ghz
where P is the maximum output
power rating of the transmitter in
watts (W) according to the
transmitter manufacturer and d is
the recommended separation
distance in meters (m). Field
strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey:
a. should be less than the
compliance level in each frequency
range.
b. Interference may occur in the
vicinity of equipment marked with
the following symbol:
)
P.A.L
d=1.2sqrt(P)
0.12
0.38
1.2
3.8
12
.
d=2.3sqrt(P)
0.23
0.73
2.3
7.3
23