Specifications - HP 54721A User Reference

Amplifier plug-in
Table of Contents

Advertisement

Specifications
Channel
Internal Trigger
External Trigger
This chapter lists the system specifications and characteristics of the
HP 54721A plug-in when it is combined with either the HP 54720 or
HP 54710 mainframes. The specifications and characteristics for the
mainframe are in the
HP 54720 and HP 54710 Oscilloscope User's
Reference.
Specifications
The following are specifications used to test the HP 6'^-721A plug-in.
Specifications are valid after a 20 minute warm-up period. See the HP
54701A Active Probe service guide for complete probe specifications.
Analog Bandwidth (-3 dB) dc to >1.1 GHz
Bandwidtli (-3 dB) with HP 54701A Active probe dc to >1 GHz
dc Voltage Measurement Accuracy (Best Accuracy calibration only)^
Single Marker ±(gain accuracy + offset accuracy)
Dual Marker ±[gain accuracy -f (2 x resolution)]
Input Resistance 50 0 ±1.5%
Trigger Sensitivity ^
High Sensitivity 0.1 div at 100 MHz; 0.3 div at 1 GHz
Normal Sensitivity 0.5 div at 100 MHz; 1.5 div at 1 GHz
Noise Reject 1 div at 100 MHz; 3.0 div at 1 GHz
Minimum Pulse Width at Normal Sensitivity ^ <500 ps at >1 div'
Trigger Sensitivity
High Sensitivity 6 mVp.p at 100 MHz; 20 mVp.p at 1 GHz
Normal Sensitivity 30 mVp.p at 100 MHz; 90 mVp-p at 1 GHz
Noise Reject 60 mVp.p at 100 MHz; 160 mVp.p at 1 GHz
Minimum Pulse Width at Normal Sensitivity ^500 ps at ^60 mVp.p