Specifications; Test Equipment Required; Frequency Response Test - HP 465A Operating And Service Manual

Table of Contents

Advertisement

V
Section
Table 5-1
Instrument
DC Voltmeter
Test Oscillator
Oscilloscope
AC Voltmeter
Variable Voltage Line
Transformer
Ohmmeter
Frequency Response
Test Set
Distortion Analyzer
DC Power Supply
Resistor
5-0
Table 5-1. Test Equipment Required
Critical Specifications
Accuracy: ±2%
Voltage R ange: 50 v full scale
Frequency Range: 10 eps to
1 Mc
].
Voltage Output:
v
Frequency: 2 kc
Vertical Sensitivity: 10 v/cm
Accuracy: ±1%, 100 cps to
50 kc
±2%, 10 cps to
1 Mc
Voltage Range: 10 v
Voltage Range: 103.5
to
126.5 v
Output
Power: 10 watts
Obms Range: 10 M obms
Frequency: 10 cps to 1 Mc
with external oscillator
Voltage Output: 10 v
Frequency: 10 cps to 500 kc
Sensitivity: 1% full scale
Voltage Output:
45 v
+
Current Limit: 75 ma
1 M ohm, 1%, 1/2 w
50 ohms, 1%, 1 w
I M ohm, shielded load
Use
Performance Checks
-hp- Model 3440A/3445A
Performance Checks
-hp- Model 651A Test
Troubleshooting
-hp- Model 130C
Performance Checks
- hp- Model 3440A/3445A
-hp- Model 33. 1 A
Superior Electric Co.
Performance Checks
Troubleshooting
-hp- Model 3440A/3445A
Performance Checks
-hp- Model 739AR Fre-
Performance Checks
-hp- Model 331A
-hp- Model 723A DC
Troubleshooting
-hp- Part No. 0757-0059
Performance Checks
-hp- Part No. 0757-0024
See Figure 5-3
Model 465A
Recommended Model
AC-DC Digital Voltmeter
Oscillator
Oscilloscope
AC-DC Digital Voltmeter
Distortion A nalyzer
Type 3PN-1l6
Multi- Function Unit
quency Response Test
Set
Distortion Analyzer
Power Supply
01775-1