Note Install the FLIR Thermography software on your computer before you
move the images.
2.12 Screen elements
See Figure 1.4 Screen elements, page 2.
1.
Main menu toolbar.
2.
Submenu toolbar.
3.
Spotmeter.
4.
Result table.
5.
Status icons.
6.
Temperature scale.
2.13 To keep in mind
By default, most cameras adapt the scale automatically. Use this mode first,
but do not hesitate to set the scale manually.
A thermal camera has a resolution limit. This depends on the size of the de-
tector, the lens, and the distance to the target. Use the center of the spot tool
as a guide to the minimum possible object size, and get closer if necessary.
Make sure to stay away from dangerous areas and live electrical components.
Be careful when holding the camera perpendicular to the target—you may be-
come the main source of reflection.
Select a zone of high emissivity, e.g., an area with a matte surface, to perform
a measurement.
Blank objects, i.e. those with low emissivities, may appear warm or cold in the
camera, as they mainly reflect the environment.
Avoid direct sunlight on the details that you are inspecting.
Various types of faults, e.g., those in a building's construction, may result in
the same type of thermal pattern.
Correctly analyzing an infrared image requires professional knowledge about
the application.