CAIRE 10848321 Service Manual page 14

Breeze oxygen concentrator

Advertisement

VII
ATF Module Replacement
98.
90.
10.
11.
ATF MODULE KIT INSTRUCTIONS
Make sure unit is turned off and unplugged.
Use a Phillips screwdriver to remove the bolts (4 typ.) for the rear cover.
Puil rear cover off.
Use a Phillips screwdriver to remove the ATF clamp clips (3 typ.).
Use a 9/32" socket to remove the hose clamp for the air intake line at the base of
the ATF module.
Slide hose off stub.
'
Pull off oxygen output line at top of the ATF module.
Remove ATF wires from wire clip near base of the module.
Remove ATF module from unit by lifting straight up and then out.
Note the
installed orientation of the ATF module with the unit.
Unplug ATF wire harness by depressing the side of the plug and then gently pull
out.
-
Disconnect spaded terminal connections (2) on older style Breeze units, which
connect directly to terminal block.
Reverse steps #9 thru #1 to install new ATF sieve module.
Note, the ATF
module has a bracket on the underside of the top dome (90° from oxygen outlet)
that clips into a support on the main unit.
Also, when
installing the ATF
module,
be alert to wire
position, so they do not get pinched.
Reinsert
wires into clip.
Plug in unit and run to verify performance.