Download Print this page

CAIRE 10848321 Service Manual page 7

Breeze oxygen concentrator

Advertisement

ll!
Preventative Maintenance Schedule
CAIRE Ref. No.:
10961833
Model:
Breeze Messenger Oxygen Concentrator
Indicated By
Weekly
As
120 Days |
365 Days
Messenger
(by patient)
|
Needed
(approx.) |
(approx.)
Reporter
Clean Cabinet Intake Filter
(Located right side of control
panel.)
Replace Cabinet Intake Filter
Check O2 Concentration at Max.
Prescribed Flowrate
+
1-5 LPM min. 90%
+
5-6 LPM min. 87%
Check Alarm Function to Verify
Adequate 9v Battery Voltage
When the unit has an oxygen concentration leve! below the minimum specification, the
following maintenance items should be reviewed:
+
Replace the compressor intake Hepa filter.
+
Replace the final filter.
+
Replace the 9 voit battery.
+
If necessary, rebuild or replace the compressor.
+
As needed, vacuum dust and dirt from the interior of cabinet.
*
Verify accuracy of Messenger.