V
Electrical System Components
CIRCUIT BREAKER
(Reference #5)
The AC electrical system is protected from overloads and short circuits by a single
pole circuit breaker. The circuit breaker is a push button reset type and the push
button is located at the top right of the contro! panel. The push button must be in
the depressed position for the Breeze unit to operate.
PRESSURE SWITCH (Reference #52)
The pressure switch is a sensor that will close a circuit activating the alarm
function when the oxygen product pressure falls below normal. This alarm circuit
has a 9 volt battery power supply and will activate even when the unit is not
plugged into an AC service outlet. For applications at higher altitudes the pressure
switch may need to be adjusted to a slightly lower set point. Call CAIREº for
assistance.
.
ATF DRIVE MOTOR
(Reference bottom of # 40)
The ATF drive motor is simply a very accurate clock drive motor that rotates a
valve that cycles the sieve beds of the ATF module.
CAPACITOR
MOTOR RUN
(Reference #30)
The air compressor motor requires a running capacitor for proper operation.
COLD START CAPACITOR
KIT (optional)
(Reference #43)
The cold start capacitor kit is a device that installs parallel into the capacitor circuit
of the compressor.
This device provides a momentary additional capacitance to the
air compressor motor during start up to provide additional torque.
This additional
~
torque will allow the unit to start at lower temperatures.
AUDIO ALARM
(Base models only).
(Reference #59)
A DC voltage audio device for the alarm function.
This function is provided for by a
device on board the OCS! circuit board for the units equipped with the oxygen
monitoring system.
Page 10