Power Supply
Charging the battery pack
AC adapter (optional)
Ni-Cd Battery Pack (optional)
AA Cells
Automatic Power Off:
Display Backlighting
Warm-up Time
Power Consumption
Insulation Resistance
Withstanding Voltage
Operating Temperature/Humidity Ranges
Storage Temperature/Humidity Range
I/O terminals
Operating altitude

External Dimensions

Weight
External Dimensions
63
AC adapter, Ni-Cd battery pack, or commercially available AA alkaline cells; the Ni-Cd
battery pack can be recharged only within the instrument.
Timer charging by the instrument (approximately 10 hours)
Input: 100 to 240 V AC, 50/60 Hz, 40 to 55 VA
Permitted Input Voltage Range: 90 to 264 V AC
Permitted Input Frequency Range: 48 to 62 Hz
1200 mAh, 9.6 V, recycle-designated part, service life of 2 years (varies according to the
operating conditions), part number B9914PS
Continuous operating time for guidance:
With generating 5 V DC (with 500 Ω load), and operating measuring function:
Approx. 7.5 hours
With generating 20 mA (with 1 kΩ load), operating measuring function, 24 V DC
power supply, EL lit: Approx. 2.5 hours
Eight AA alkaline cells
Continuous operating time for guidance:
With generating 5 V DC (with 500 Ω load), and operating measuring function:
Approx. 10 hours
With generating 20 mA (with 1 kΩ load), operating measuring function, 24 V DC
power supply, EL lit: Approx. 2 hours
Approximately 30 minutes under the following conditions: No key or communication
operation is performed while being run on batteries.
Segmented LCD, EL backlighting
Approx. 5 minutes
55 VA or less
20 MΩ or more for one minute at 500 V DC (between terminals/each terminal and AC
adapter power line)
350 V AC for one minute (between terminals)
1500 V AC for one minute (between each terminal and AC power line)
For generation or measurement (normal), or charging without generation or measurement
running:
5 to 40°C, 20 to 80% RH (no condensation)
For charging with generation or measurement running:
5 to 30°C, 20 to 80% RH (no condensation)
-20 to 45°C, 90% RH or less (No condensation allowed)
Maximum applied voltage is less than 42 Vpeak and CAT II between terminals/each
terminal and ground. Except, negative 24 VOUT terminal is 18 Vpeak or less.
Max. 2000 m
237 x 137 x 63 mm (does not include projections)
Approx. 1.2 kg (including 8 dry cells)
237
Unit:mm
55