3. Se�ngs menu „SET"
Protected by password 0004. By entering a wrong password, the detector returns to measurement mode.
The password can be changed in the menu SET -> PSWD.
Menu
Calibra�on
Se�ng the alarms
Se�ng the range of the 20 mA
current loop
Changing the password
Se�ng the year
Se�ing the date
Se�ng the �me
Exi t
Calibra�on
a) Connect fresh air to the sensor input. The icon
"
b) Using „
" select the concentra�on of the calibra�on gas, confirm „
c) Connect the calibra�on gas to the sensor input. The icon „ " flashes. Wait un�l the value stabilizes and confirm
"
d) Using „
" swtch between „YES" - save calibra�on, or „NO" - return to the Se�ngs menu. Confirm the selected
op�on „
"
e) In case of saving a new calibra�on, the detector will restart.
Se�ng the alarms
a) Using „
" select the concentra�on for alarm 1. Confirm „
b) Using „
" select the concentra�on for alarm 2. Confirm „
c) Using „
" select the concentra�on for alarm 3. Confirm „
Display
"
Descrip�on
Sensor calibra�on
Se�ng the alar ms
Se�ng the range of the 20 mA
current loop
Changing the password
Se�ng the year
Se�ng the date
Se�ng the �me
flashes
A�er the value stabilizes, move onto the next step using
.
"
"
"
"