  • Page 1 Quick Start Guide for mass flow meter & controller FLOW IQF-100C, IQF-200C IQFD-100C, IQFD-200C IMPORTANT! Read carefully before use. Keep for future reference.
  • Page 2: Safety First

    ▶ If in any doubt: scan the QR code (Service & Support) on the The product requires professional experience and knowledge of back cover to get in contact with Bronkhorst. installation, operational requirements of the relevant industry, and knowledge of the type of application and national requirements ▶...
  • Page 3: Process Fluid

    ▶ If in any doubt: scan the QR code (Service & Support) on the product in a non-critical process. It guides the user in controlling back cover to get in contact with Bronkhorst. and monitoring research and pilot setups by means of The user is responsible for taking the necessary safety measures Bronkhorst FlowSuite.
  • Page 4 FLOW - IQF-x 1. Fluid inlet 5. Fluid outlet 2. Serial number label 3. Switch/Push button 4. Connector / status LEDs...
  • Page 5 FLOW - IQFD-x 1. Fluid inlet 5. Fluid outlet 2. Serial number label 3. Switch/Push button 4. Connector / status LEDs...
  • Page 6: Before Installation

    5. Do not connect ▶ Scan the QR code (Service & Support) on the back cover. 6. TX-RS232/BUS ▶ Go to the Bronkhorst website and select your operating 7. +Us (15 ... 24 Vdc) system to download Bronkhorst FlowSuite. For download 8.
  • Page 7: Pressure Rating

    Mounting location (general) To ensure reliable measurements, it is important to choose a correct mounting location. For optimal performance, observe the following guidelines: ▶ Install the product on a flat, rigid, and clean surface [1]. ▶ Avoid installation close to mechanical vibrations [2]. ▶...
  • Page 8 Mounting (downported) CAUTION Risk of personal injury due to fluid leakage! Incorrect mounting can lead to fluid leakage and severe injury. ▶ Only use supplied seals and bolts for mounting. ▶ Check all connections before and after applying fluid pressure. The product is supplied with bolts and pre-mounted seals.
  • Page 9: Electrical Connection

    Fluid line connection Electrical connection For basic setup, it is recommended to use Bronkhorst® CAUTION accessories (not included) as described in this manual. Risk of personal injury due to fluid leakage! ▶ Connect the product to a RJ45 modular Y-adapter.
  • Page 10: Adjusting Zero Point

    Adding the product to Bronkhorst FlowSuite When Bronkhorst FlowSuite is started, the software will automatically search for connected instruments. All found instruments are displayed in the left pane. ▶ Use the refresh option to check for newly connected products. If the product is not found, it may not be set to serial RS-232.
  • Page 11 Product status in Bronkhorst FlowSuite If a connection between the product and Bronkhorst FlowSuite In Bronkhorst FlowSuite, an icon is displayed next to the product cannot be established, the switch/push button can be used to in the left pane. The icon indicates the status of the product. When...
  • Page 12: Maintenance

    Maintenance Technical specifications Although no routine maintenance is required, to ensure that the Description Unit Value product works correctly and safely, a regular (visual) inspection is Maximum fluid pressure bar(g) recommended. Check prior to each use Permissible fluid temp. °C (°F) 5 ...
  • Page 13: Service And Support

    (WEEE). Bronkhorst representative. Bronkhorst will take care of proper dismantling, recycling ▶ To help you quickly, make sure you have the serial number of and/or reuse. By ensuring this product is disposed of the product at hand.
  • Page 14: Declaration Of Conformity

    The full text of the applicable declarations of conformity are available at the following web address: https://www.bronkhorst.com/downloads. © 2024 Bronkhorst High-Tech B.V. - All rights reserved Bronkhorst® is a registered trademark. All trademarks are the property of their respective owners.
  • Page 16 Bronkhorst High-Tech B.V. 31 573 45 88 00 Nijverheidsstraat 1a info@bronkhorst.com NL-7261 AK Ruurlo, The Netherlands www.bronkhorst.com Service & Support Bronkhorst FlowSuite Product Info 9.17.177A • 2024-01...