3.5 Cleaning of Cell Holder in Sample Compartment
If sample is spilt in the sample compartment, clean the inside of the compartment if necessary by removing
the cell holder and cell holder base. Following is the procedure for this.
(1)
Open the top cover of the sample compartment. Figure 3-19 shows a top view of the sample
compartment interior.
(2)
To remove the cell holder, remove the holder retaining screw (one).
(3)
To remove the cell holder base, remove the three holder base retaining screws. Avoid loosening
any other screws.
Cell holder base
retaining screws
(three)
Fig. 3-19
3 - 22
Cell holder
Cell holder retaining
screw (one)
Cell holder base