3-5 Local G Value Setting
Set switch SWA1 the main board to ADJ position. Press
scale, and press
ON|ZERO
f
Enter the local G value
Press
key
TARE
The display shows and
saves the entered G value
in EEPROM
Set switch SWA1 to the LOCK position.
The scale returns to weighing mode.
The G value should be among the value of Equator and Polar.
Acceleration of gravity at the Equator: G
Acceleration of gravity at the Poles: G
Taipei ≒ 9.78914 m/sec
When the G value is NOT within the range 9.772~9.829, the display shows error
massage "
02025100
f
key again before the display resets back to zero.
.
.
.
P
2
Shanghai ≒ 9.79423 m/sec
".
ON|ZERO
f
Display current G value
ON|ZERO key Upward key (from 0 to 9)
key Downward key (from 9 to 0)
UNITS
1
TARE key Confirm / enter key
NET|GROSS key Move the cursor to left.
= 9.7803184558 m/sec
E
= 9.8321772792 m/sec