Page 2
1. ON/OFF Button 2. Display 3. M Button (Memory) 4. Bluetooth/Time Button 5. User Button 6. Cuff Socket 7. MAM Switch 8. Traffic LED Light indicator 9. USB Port 10. Mains Adapter Socket 11. Battery Compartment 12. Cuff 13. Cuff tube 14.
Page 3
Indications for Use The Upper Arm Blood Pressure Monitor, Model BP3KV1-5K is a device intended to measure the systolic and diastolic blood pressure and pulse rate of an adult individual by using a non-invasive oscillometric technique in which an inflatable cuff is wrapped around the upper arm for a circumference range from 22 to 42cm.
Page 4
When replacing the batteries, ensure all the batteries are replaced simultaneously to avoid battery damage and potential damage to the monitor. Microlife USA does not recommend using rechargeable batteries. To avoid inaccurate measurements and to lessen any discomfort from Cuff pressure, ensure the Cuff is placed correctly on the limb and fits correctly when snug (not tight), as indicated by markings with the Cuff.
Page 5
Use third party accessories. Only use Microlife authorized accessories, such as cuffs or AC adapters, as those not approved for use with the device may provide inaccurate measurements, injury, and/or damage the device.
Page 6
Changes or modifications to the product are not approved by Microlife USA and could void the user's authority to operate the equipment under FCC jurisdiction. This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to Part 15 of the FCC Rules.
• Which batteries and which procedure? • Using rechargeable batteries 7. Using a Mains Adapter 8. Bluetooth® Function • Connecting the «Microlife Connected Health+» App 9. PC-Link Functions 10. Error Messages 11. Safety, Care, Accuracy Test and Disposal • Safety and protection •...
Now you can enter the new values as described above Selecting the correct cuff Microlife offers different cuff sizes. Select the cuff size to match the circumference of your upper arms (measured by close fitting in the centre of the upper arm). Cuff size for circumference of upper arm 17 –...
Page 9
reliable than a single measurement is performed. •When you select MAM mode, the MAM-symbol 24 appears in the display. •To select standard mode, slide the MAM switch 7 on the side of the device downwards to position «1» and to select MAM mode, slide this switch upwards to position «3».
3. Taking a Blood Pressure Measurement 1. Select standard (single measurement) or MAM (Automatic double measurements); see details in chapter «1». 2. Press the ON/OFF button 1, the backlight will be turn on and to start the measurement. 3. The cuff will now pump up automatically. Relax, do not move and do not tense your arm muscles until the measurement result is displayed.
following range defined by ACC, AHA 2017 Guidelines. Data in mmHg. Range Systoli Diastolic Recommendation blood pressure too ≧130 ≧80 Seek medical advice high blood pressure <80 Self-check elevated -130 blood pressure <120 <80 Self-check optimum The higher value is the one that determines the evaluation. Example: a blood pressure value of 140/80 mmHg or a value of 130/90 mmHg indicates «blood pressure too high».
Information for the doctor on frequent appearance of the irregular heartbeat symbol This instrument is an oscillometric blood pressure monitor device that also analyzes pulse frequency during measurement. The instrument is clinically tested. If pulse irregularities occur during measurement, the irregular heartbeat symbol is displayed after the measurement.
display first shows« M » 25 and a value, e.g. « M17 ».single values in memory. Pressing the M-button again displays the previous value. Pressing the M-button repeatedly enables you to move from on stored value to another. * Pay attention that the maximum memory capacity of 99 memories is not exceeded. When the 99 memory is full, the oldest value is automatically overwritten with the 100 value.
7. Using a Mains Adapter You can operate this device using the Microlife mains adapter (DC 6V, 600 mA). * Only use the Microlife mains adapter available as an original accessory appropriate...
- Automatically turn off Bluetooth: Bluetooth will turn off automatically after 2 minutes if a smartphone does not connect to the device. Bluetooth pairing & App setup: 1. Open Microlife Connected Health+ App on the smartphone. (Make sure the App is running on the foreground, not in the background.)
9. PC-Link Functions This device can be used in conjunction with a personal computer (PC) running the Microlife Blood Pressure Analyser (BPA+) software. The memory data can be transferred to the PC by connecting the monitor via a cable. If no download-voucher and cable is included download the BPA software from www.microlife.com/technologies/connect/bpaplus...
App on the smartphone. Wait for 1 minute, open the App on the smartphone and manually activate Bluetooth on device, to re-try Bluetooth connection & data transfer. «ERR Bluetooth Bluetooth is malfunctioning. Contact your local Microlife bt» self-check error. distributor. «HI»...
* Please consult your doctor, if this or any other problem occurs repeatedly. If you think the results are unusual, please read through the information in «Section 2» carefully. 11. Safety, Care, Accuracy Test and Disposal Safety and protection • This device may only be used for the purposes described in these instructions. The manufacturer cannot be held liable for damage caused by incorrect application.
WARNING: Do not wash the cuff in a washing machine or dishwasher! Accuracy test We recommend this device is tested for accuracy every 2 years or after mechanical impact (e.g. being dropped). Please contact your local Microlife-Service to arrange the test (see foreword). Disposal Batteries and electronic devices must be disposed of in accordance with the locally applicable regulations, not with domestic waste.
Page 20
Technical alterations reserved. The Bluetooth® word mark and logos are registered trademarks owned by Bluetooth SIG, Inc. and any use of such marks by Microlife Corp. is under license. Other trademarks and trade names are those of their respective owners.
Need help?
Do you have a question about the BP3KV1-5K and is the answer not in the manual?
Questions and answers