The following table shows the specifications for replacement fuses:
For 220-240 VAC Systems
3 ampere
250 volt
5 x 20 mm
Series 3AG
F-type (also known as
quick-acting or fast-blo)
Warning
To avoid the risk of electrical shock, always turn off the GC
interface and disconnect it from the wall outlet or power strip before
changing fuses.
Warning
To avoid shock and fire hazard, always use replacement fuses of the
specified type and current rating.
Follow these steps to check and replace fuses:
1. Turn off the module power by pressing the power switch on
the rear panel, and then unplug the power cord.
Warning
To avoid shock and fire hazard, do not attempt to change the fuses
with the power cord plugged in.
2. Disconnect the module power cord from the connector on the
rear panel.
For information about rear panel connectors, see "Rear panel" in
the "Introduction" chapter.
For 100-120 VAC Systems
6 ampere
250 volt
0.25 x 1.25 in
Series 3AG
F-type (also known as
quick-acting or fast-blo)
TGA-IR Module User's Guide