Thermo Fisher Scientific
4. Connect the exhaust gas line to the gas cell exhaust gas
connector.
Gently push the Swagelok
the gas cell exhaust gas connector. Turn the Swagelok fitting
until tight to secure the connection.
5. Connect the exhaust gas line to a venting system.
Connect the exhaust gas line to a vent hood or whatever venting
system your facility uses. We also recommend using an aspirator
to draw out the exhaust gases. If you use and aspirator, set it for
10 inches Hg or less.
®
fitting on the exhaust gas line into
Gas cell exhaust
gas connector
Exhaust gas line
to venting system