PORTABLE SENSOR DISPLAY – STRAIN BRIDGE

Polynomial

This type of calibration consists of a list of values used to perform polynomial
linearization.
Figure 48. Polynomial Screen
Polynomial calibration allows linearization of up to 11 points. Note: the Points [+]
button must be used to add one point at a time. Once the polynomial value has been
entered, the button will be enabled to allow another point to be added.
,
NOTE
This toolkit does not calculate the polynomial values! It is assumed that the user is
already in possession of appropriate polynomial values to enter.
TABLE 28. POLYNOMIAL SCREEN
Item
No header
Value edit field
Description
This simply indicates which calibration point the row
refers to.
Enter the polynomial value for this point