Connections; Strain Bridge Sensor Field Cable Connector Wiring - Honeywell 7561-PSDS User Manual

Portable sensor display
Table of Contents

Advertisement

MODEL 7561-PSDS
PORTABLE SENSOR DISPLAY – STRAIN BRIDGE

Connections

Strain Bridge Sensor Field Cable Connector Wiring

To attach this connector to the handheld, align the
white arrow on the connector with the white line
on the handheld socket. Rotate the locking collar
as indicated on the connector. The connector
fitted to PSDS standard product is BINDER 770-8.
The cable-fitted mating connector is a BINDER
771 8-way male connector. This male connector
is available under three different BINDER order
codes determined by the field cable diameter.
TABLE 8. CONNECTOR INFORMATION
Field cable diameter
(mm)
2,5 to 4,0
4,0 to 6,0
6,0 to 8,0
Recommended cable: BELDEN 9504 4-pair cable. Max cable size 0.25 mm² , AWG 24.
List of conductor identification and functions.
TABLE 9. CONDUCTOR ID AND FUNCTION
Cable Pair
Suggested
Color
White
1
Black
Green
2
Black
Red
3
Black
Blue
4
Black
Shield/
Grey
Screen
View from solder connector side of the connector
Binder Order Code
99 0771 000 08
99 0771 001 08
99 0771 002 08
Connector
Function
Pin
1
Bridge sensor reference +ve
2
Bridge sensor reference -ve
3
Bridge sensor signal +ve
4
Bridge sensor signal -ve
5
Bridge sensor excitation +ve
6
Bridge sensor excitation -ve
7
TEDS
8
Ground
Cable screen should only be connected
to chassis of the sensor. If this cannot
be achieved, it should be connected to
Excitation –ve
Honeywell P/N
Not stocked
023-1721-00
023-1720-00