OWNER'S MANUAL (Includes Operation, Maintenance and Parts) DATA & TOUCH SCREEN WARMING CABINET MODELS STAINLESS WARMING CABINETS Read this manual before starting to work! This information is necessary for safe and efficient operation of the equipment. TEC-F-0579 REV2 01/09/2023...
Page 2
Any translations must be from the base language document. Printed copies are not controlled documents. Although current at the time of publication, SKYTRON'S policy of continuous development makes this document subject to change without notice. If a current document is required, contact your local SKYTRON representative above.
Warming Cabinet functioning at peak efficiency. CAUTION The design of the SKYTRON warming cabinet Indicates a hazardous situation that, if not avoided, does not utilize internal user serviceable parts could result in minor or moderate injury.
Page 6
For optimal product memory to malfunction. usage, safety and durability, service must Any parts or assemblies not listed in this be performed by a SKYTRON authorized service technician using SKYTRON section must be serviced or replaced by SKYTRON authorized service personnel only.
Page 5 SECTION 2. EQUIPMENT SPECIFICATIONS 2-1. Intended Use 2-2. Recommended Settings SKYTRON does not recommend chamber set points for any items that are to be warmed. For appropriate CAUTION heating temperatures, please contact the item manufacturers. For more information, please Explosion or Fire Hazard: The Warming contact your SKYTRON service representative.
MANUFACTURER'S RECOMMENDATIONS FOR SET POINT Toll Free (800) 759-8766 • Press the power switch to the ON position 1" OF www.skytron.us FOR IRRIGATION FLUIDS FOLLOW SPACE TEMPERATURE GUIDELINES PRINTED ON 2. To change the setpoint: CONTAINER OR CONTACT YOUR SUPPLIER...
Page 9
Page 7 2 & 7 ON THE SHELVES SURFACE CHAUDE SURFACE CHAUDE Contact avec la peau Contact avec la peau peut causer des brûlures peut causer des brûlures NE PAS TOUCHER NE PAS TOUCHER H0800-08 H0800-08 Cet équipement HAZARDOUS VOLTAGE This Equipment ENCLOSED doit être...
The power cord is located at the rear of the cabinet in the upper left hand corner. SKYTRON offers many models of Warming The main power ON/OFF switch/circuit breaker is Cabinets to meet each facility’s unique needs: located on the front control panel next to the display.
Loading any item that could introduce flammable agents into the cabinet SKYTRON recommends personal protective atmosphere can cause a fire or explosion. equipment while loading or unloading contents. 1. Make sure power cord is plugged into a properly Load only sterile water, saline solutions, or (Muslin, grounded 120VAC electrical outlet.
Page 12 Contents should be loaded to allow for a minimum of one inch from the top, back and sides of the WARNING compartment. Minimum spacing of one quarter Injury or Burn Hazard: DO NOT use inch between fluid containers is recommended for liquids on or inject in living tissue unless evenly distributed heating (Figure 6).
Page 13 SECTION 4. CONTROLS 4-1. Standard Controls The controls are located on the upper panel on the front of the cabinet. For dual-chamber cabinets there are individual controls for each chamber. Each control is clearly labeled UPPER and LOWER for the dual chamber cabinet (Figure 8).
AORN guidelines and the fluid manufacturer’s may be down loaded to a USB drive at any time. recommendations for proper temperature settings. 1. Insert the SKYTRON flash drive (W0600-S) or equivalent into the USB port marked DATA, on the front face of the control panel.
Page 15 4-2. Touch Screen Controls The touch screen is located in the center of the upper panel at the front of the cabinet. A single touch screen controls single or dual chamber cabinets (Figure 11). WARNING The USB port should only be used for downloading temperature history data. DO NOT connect USB wireless devices due to potential EMI conditions.
Page 17 4-2-5. Notifications WARNING Current settings, operational conditions and the current logged in user can be quickly viewed. Explosion Hazard: DO NOT exceed 150° F (66 °C) for non-vented closures; screw Touch: (1) yellow time/date to open a drop-down caps, crimp seals, plastic pouches, etc.
Page 18 4-2-7. Data File 3. Touch: (6)ID#1 or #2 for added file identification. If: (5)Fixed Interval is ON, interval is based on − Data File Naming Length(days) with 1 day equal to a 1 minute 1. Touch: (1) Home (2) Monitor (3) swipe screen interval, 7 days, the logging rate is every 7 minutes.
Page 21
Page 19 Select a button on the screen. MM/DD/YYYY HH:MM AMPM Axis 1 Return Time Period (minutes) Left Right Chart Maximum Maximum Figure 29. Chart Icon Screen Minimum Minimum -100 Automatic Scale Automatic Scale 2. At Chart screen, touch (1)Labels to verify color and values (2)Scroll arrows cycle through each Figure 32.
Page 22
Page 20 CAUTION: NOTICE Once files are deleted, they are gone and Signature screen is not available if security is can not be retrieved. Do Not edit, move, or disabled. System security must be enabled delete other files, doing so can cause internal to digitally sign data log files.
Page 21 MM/DD/YYYY MM/DD/YYYY MM/DD/YYYY HH:MM AMPM HH:MM AMPM HH:MM AM PM Signature Alarm File Open Verify Save Data file is not active ALARM MESSAGE MM/DD/YYYY HH:MM AMPM MM/DD/YYYY HH:MM AM PM User: First Name Last Name Select Email Address ALARM MESSAGE First LastName Select...
It is recommended to perform preventative to the grain. maintenance on the SKYTRON Warming Cabinets 5-2. Sterilization every six months or when any service is performed. WARNING WARNING Explosion Hazard: DO NOT exceed 150°...
The specific items listed in the Warming Cabinet 5-5. End Of Useful Life And Disposal Maintenance Matrix shall be inspected and repaired The end of the useful life for the SKYTRON Warming or replaced as necessary. Cabinet is 10 years under normal operating conditions, service parts are available for this period.
Alarm Reset button until it quits. Wait for contents to cool to unload, then reload the contents. Turn the chamber on and monitor performance. If it continues to overheat to an alarm (HI) condition, turn off the chamber and contact your SKYTRON authorized service personnel.
Page 26 6-2. Touch Screen Controls Troubleshooting Guide Alarm Description Explanation/Corrective Action 1. Is the circulating fan operational? 2. Are the contents loaded properly? 3. Has the chamber set temperature been lowered? 4. Lower chamber temperature cannot be in excess of +30° F (+1.1 °C) above an upper chamber. When cabinet temperature exceeds the setpoint by 10°...
SECTION 7. REPLACEMENT PARTS Warming Cabinet replacement parts listed in this CAUTION section have been identified by SKYTRON as Any parts or assemblies not listed in this serviceable by facility personnel and are available section must be serviced or replaced by for purchase.
Page 31
Page 29 7-1. SINGLE AND DUAL CHAMBER REPLACEMENT PARTS Item Part No. Description Qty. W0102-KEY Key Only - Key Lockout, two (2) keys per set ..........AR W0600-S USB Drive, Data Retrieval ................AR W0050 HANDLE, Stainless Steel Door, LH Upper, RH Lower ........1 W0050-CH HANDLE, CuVerro antimicrobial, Steel Door, Dual Chamber, Upper....1 W0053...
Revision History 07/10/2018 Initial release. 03/27/2020 Reformatted document to current Skytron standards. Revised "Intended Use" on page 5. Added section "Recommended Settings" on page 5. Added section "Maximum Warming Temperature Limit for Patient Safety" on page 12. Added USB warning to section 4-1 "Standard Controls" on page 13 and section 4-2 "Touch Screen Controls"...