Cable gland
• Clamping range:
- 5 to 10 mm (0.2 to 0.39 in) acc. to EN 50262 or 7 to 10 mm (0.28 to 0.39 in) acc. to
UL-514 B (Cable gland made of plastic)
- 7 to 10.5 mm (0.28 to 0.41 in) (Cable gland made of metal)
• Tightening torque
- Max. 6 Nm (Cable gland made of plastic)
- Max. 10 Nm (Cable gland made of metal)
For the optionally supplied mating M12 connectors, the following requirements apply:
• Clamping range of the cable: 6 to 8 mm (0.24 to 0.31 in)
• Cable cross-sections: Max. 0.75 mm
Suitable connection cable
→ 41 and order structure option "accessory enclosed"
6�2

Connect potential equalization

The potential equalization for the device must be integrated into the existing potential
equalization on site.
Requirements:
• The potential equalization must be connected to the external ground terminal on the
device.
• For optimum electromagnetic compatibility, keep the potential equalization line as short
as possible.
• The recommended cable cross-section is 2.5 mm
• The potential equalization of the FDR56/FQR56 must be included in the local potential
equalization.
6�3

Connecting the device

The electrical connection is realized by internal terminals or external connectors.
Insert cable into cable gland.
- For plastic cable gland: Tighten cap nut until the rubber seal is touched all around
and then tighten cap nut with by ½ turn.
- For metal cable gland: Tighten the cap nut (torque of up to 10 Nm).
Connect the protective ground.
Connect the power supply and signal output.
For connection via plug connector:
Connect the protective ground.
Connect the connectors and fix them.