Clearing The Memory; Battery Operation - Nutricia Flocare Infinity Instructions For Use Manual

Hide thumbs Also See for Flocare Infinity:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 86
EN
All parameters and values can be cleared one after the other in the following way:
• Press "START/STOP" to pause the pump.
• Select the parameter that needs to be cleared.
• Press the "CLR" key to clear the memory. The parameter or value returns to its default value:
ml/h = 0 ml/h
VOL = cont
= No dose set. Pump will feed continuously until feed container is empty or pump
INFO = 0 ml
= Volume delivered is cleared
In order to keep a clear overview of the daily amount of feed delivered, clear the total volume delivered as each daily feeding
schedule is started, as follows:
START/STOP
CLR
• When another feeding program needs to be started, or the pump will be used for another patient, all settings (rate and dose)
can be returned to default and the volume delivered can be set at zero by simply pressing the "CLR" key and keeping it pres-
sed for 2 seconds.
• The Flocare® Infinity™ pump is equipped with a Lithium-ion battery. The expected battery life is approximate 2 to 5 years.
Performance may degrade due to age, excessive temperatures and number of charging cycles.
• The battery symbol in the display is automatically replaced by the plug symbol when the charger is plugged in. The "fuel
gauge" then indicates that the battery is charging by displaying the bars in an ascending low to high pattern starting with
the leftmost bar.
This pattern continuously repeats while the pump is charging.
• To check the status of the battery, disconnect the charger from the pump and turn the
pump on. The bars between E and F (E=Empty, F=Full) represent the "fuel gauge" of the
battery. Each bar is approximately 1/4 of a full battery charge. If 2 bars appear the
battery is half full and remaining operating time is approximately 12 hours at a flow rate
of 125 ml/h for a new battery.
• In case of power failure, when the pump is connected to mains, the pump automatically switches to battery powered
operation.
• When during functioning the last bar is gone, the battery symbol will blink to indicate that there is approximately 1 hour of charge left.
The display will flash "BATT" every 3 seconds, alternating with the active display and the pump will beep every 2 seconds to remind
the user of the low battery charge condition. Plug in the charger to continue to run and recharge the battery.
• In case of battery failure, caused by, for example, an excessive temperature condition (which may resolve itself with time) or by a
complete battery failure, the E, F and battery symbol will flash. Turn the pump "off" and contact your healthcare provider or refer
the pump to service.
403259772_r1.indd 91
403259772_r1.indd 91

CLEARING THE MEMORY

is stopped
• Switch the pump "ON"
• Immediately after the self-test the pump displays total volume delivered.
• Press the "CLR" key. The total volume delivered returns to "0 ml".
• The pump displays the previous programmed flow rate (ml/h).
• The pump is in the "HOLD" mode ready for programming.

BATTERY OPERATION

-91-
18/04/23 7:03 PM
18/04/23 7:03 PM