Programming The Pump - Nutricia Flocare Infinity Instructions For Use Manual

Hide thumbs Also See for Flocare Infinity:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 86
EN
appears in the display. The pump will automatically stop when the Flocare
feed. However this "FILL SET" maneuver can at anytime be stopped by pushing the "FILL SET" key a second time. The pump
will return to the hold mode when the "FILL SET" maneuver is complete or stopped.
The pump has a liquid crystal display (L.C.D.) with large alphanumeric characters, smaller symbols, words and a back light.
The following information can be found in the display:
• Flow rate (ml/h), volume (ml), is displayed through the large characters.
Words below describe what the number relates to (rate, dose or volume).
The pump also displays messages, for example "end of dose" will appear when the pump has finished
delivering a single feed dose.
Figure 6
The back light of the display turns off 10 seconds after the last key is pressed.
The back light will switch on for 10 seconds when the charger is connected to the mains.
• The Flocare® Infinity™ pump can be used for continuous or dose feeding.
• Insert the pump set in the pump, switch the pump on and fill the pump set when necessary (see paragraph: filling the pump
set).
• The display shows the total volume delivered since the pump was last cleared.
• If necessary clear the volume delivered by pressing the "CLR" key.
• The last installed flow rate (ml/h) is now displayed.
• The pump is now in the "hold" mode and is ready for programming.
• We strongly recommend that for those patients where interruptions or delay in therapy could affect their health status, the
DOSE setting of the Infinity pump must be used and programmed.
403259772_r1.indd 89
403259772_r1.indd 89
Pack Infinity
®
DISPLAY
• The arcs around the run symbol rotate when the pump is running.
• The remaining battery capacity (see paragraph: battery
operation).
• The wall plug symbol indicates that the charger is plugged in.

PROGRAMMING THE PUMP

-89-
pump set is completely filled with
18/04/23 7:03 PM
18/04/23 7:03 PM

Advertisement

Table of Contents
loading

This manual is also suitable for:

35682

Table of Contents