Nutricia Flocare Infinity - Pump Troubleshooting

Why your pump alarms and how to stop it

  1. Read the screen 
  2. Press the START/STOP button to silence the alarm
  3. See overleaf for a list of reasons why your pump is alarming and how to deal with each alarm
ALARM REASON SOLUTION
The set is not fitted or is fitted incorrectly Reinsert the giving set as per instructions
The pressure sensor is dirty Clean the sensors and reinsert cassette
Door is broken, cracked or not closed Make sure the door is closed Tape door to pump tightly to continue feeding and call your Nutricia Homeward Customer Service team to arrange for a replacement
Flow rate has not been programmed Make sure the correct flow rate is programmed
The pump has been paused for three or more minutes By pressing the START/STOP button to silence the alarm, a further three minute pause commences. When ready to restart feeding press the START/ STOP button
Air bubble in the giving set Disconnect the giving set from the feeding tube and hold over a clean container. Press the FILL SET button until the air bubble has passed out of the giving set. Press FILL SET to stop, reattach to feeding tube and press START
The bag of feed is empty If further feeding is required, replace the empty bag of feed and recommence feeding. If you need to refill the giving set with feed, follow solution instructions for "air bubble in the giving set". If required dose delivered, switch off pump and discard feed and giving set as per training
The air sensor is dirty Clean the sensor area following the cleaning guidelines
The pump has administered the dose Press and hold the ON/OFF button until it beeps and is powered off
If the END OF DOSE alarms inappropriately contact your Nutricia Homeward Enteral Nurse Specialist or the Out of Hours Advice Line
The pump detected an occlusion between the feed and the pump Make sure there are no kinks in the tubing, that clamps are not closed between the feed container and the pump, and that the foil has been pierced
The sensor is dirty Clean the sensor area (see cleaning guidelines), reinsert the giving set in the pump and restart the pump
The pump detected an occlusion between the pump and the patient If relevant, make sure the clamp is open
Make sure there are no kinks in the giving set or feeding tube
The sensor is dirty Clean the sensor area (see cleaning guidelines), reinsert the giving set in the pump and restart the pump


Do not use the 'FILL SET' function when there is a set occlusion as this may cause the pump to incorrectly calibrate

HOW TO STOP THE OCC ALARM

HOW TO STOP THE OCC ALARM

  1. Press the START/STOP button 
  2. Detach the giving set
  3. Solve the occlusion problem (see below) 
  4. Re-insert the giving set and close door 
  5. Press the START/STOP button

THE MOST COMMON SET OCCLUSION

THE MOST COMMON SET OCCLUSION

ALARM REASON SOLUTION
The battery capacity is running low. There is approximately one hour of feeding left when a rate of 125ml/hour is set Connect the pump to the electrical supply and charge for 6 hours. The pump can be used while connected to the electrical supply
The battery has failed Contact your Nutricia Homeward Customer Service team to arrange for the pump to be serviced
The current feeding regimen has been locked into the machine to prevent accidental changing of regimen Contact your Nutricia Homeward Enteral Nurse Specialist

ERROR with a number, e.g.
ERROR 12
The self test detected a fault Contact your Nutricia Homeward Enteral Nurse Specialist to arrange for the pump to be serviced

Nutricia Advanced Medical Nutrition White Horse Business Park, Trowbridge, Wiltshire BA14 0XQ
www.nutriciaflocare.com
SCC3297-08/17

Documents / Resources

References

Download manual

Here you can download full pdf version of manual, it may contain additional safety instructions, warranty information, FCC rules, etc.

Download Nutricia Flocare Infinity - Pump Troubleshooting

Advertisement

Need help?

Need help?

Do you have a question about the Flocare Infinity and is the answer not in the manual?

Questions and answers

Table of Contents