USER MANUAL
CRYOLYS
Evolution
®
2.1.
Description of the equipment
1
2
3
4
5
6
7
8
9
Page
8
1
Main seal: ensures a good seal between the cover and the
module.
2
Diffuser Plate: this is attached to the cover and directs air over
the dry ice.
3
Cooling module: ensures uniform diffusion of cooled air inside
the Precellys® Evolution.
Module cover: Encloses the cooling module.
4
Locking levers: allow removal or locking of the cooling mo-
5
dule from/to the Precellys® lid.
6
Precellys® lid: its fixed base receives the cooling module.
7
Fan: propels atmospheric air through the module. The four-
pin connector must be connected to the back of the Precellys®
Evolution.
8
Temperature sensor: measures the temperature inside
the Precellys® Evolution and detects whether the module is
attached to the Precellys® lid. The three-pin connector must be
connected to the back of the Precellys® Evolution.
9
Hinge reinforcement: reinforces the hinge of the Precellys®
Evolution in order to support the Cryolys® Evolution.
2.2.
Specifications
Weight (empty module)
1.1 kg
Operating temperature
15-30°C
Humidity
15-85% RH
Altitude
< 2,000 m
Operating specifications
Maximum air flow
1.8 kg
Dry ice reservoir
ON or OFF
Dry ice consumption
Approx. 1.5 kg/30 mins
Minimum dry ice volume
0.5 kg
Noise level
< 60dB
Temperature control
Between 0°C and 10 °C (1°C increments)
Page
9
Need help?
Do you have a question about the CRYOLYS Evolution and is the answer not in the manual?