Electrical Connection - Miele PLW 8617 Safety Operating Instructions Manual

Washer-disinfector
Hide thumbs Also See for PLW 8617 Safety:
Table of Contents

Advertisement

 All electrical work must be carried out by a suitably qualified
electrician in accordance with local and national safety regulations.
- The electrical installation must comply with current local and na-
tional safety regulations (e.g. DIN VDE 0100 in Germany).
- The connection to the power supply must be via a suitably rated
plug and socket and must comply with national regulations. The
socket must be accessible after the machine has been installed. An
electrical safety test must be carried out after installation and after
any maintenance work.
- If the machine is hard-wired to the power supply, a power switch
with all-pole isolation must be installed. The power switch must be
designed to operate at the rated current for the machine, must en-
sure a 3 mm gap between all open contacts, and must be able to
be locked in the off position.
- Equipotential bonding should be carried out if required.
- The rated loads are specified on the data plate and in the circuit
diagram supplied with the machine.
- Check that the phase rotation is correct when connecting the ma-
chine. Failure to do so could have an adverse effect on the circula-
tion pump and impair the wash quality.
- For added safety, the machine should be protected by a type B re-
sidual current device with a trip current of 30 mA (DIN VDE 0664).
You can find more information on electrical connection in the in-
stallation diagram supplied with the machine.
The machine must only be operated with the voltage, frequency and
fuse rating shown on the data plate.
The data plate showing relevant test marks is located behind the ser-
vice panel on the infeed side of the machine.
The wiring diagram is supplied with the machine.

Electrical connection

79