g) Movement error symbol (
)
The Movement Error Symbol (
during the measurement. Please remove the cuff, and wait 2-3 minutes.
Reapply the cuff and take another measurement.
Patient Position:
NOTE:
1) Comfortably seated
3) Feet flat on the floor
5) Middle of the CUFF at the level of the right atrium of the heart
5.4. Irregular Heartbeat Detector
This symbol(
) - indicates that certain pulse irregularities were detected
during the measurement.
In this case, the result may deviate from your normal basal blood pressure –
repeat the measurement.
Information for the physician
Heartbeat Symbol.
This instrument is an oscillometric blood pressure monitor device that also
tested.
If pulse irregularities occur during measurement, the irregular heartbeat
symbol is displayed after the measurement. If the symbol appears more
if it suddenly appears more often than usual, we recommend the patient to
seek medical advice. The instrument does not replace a cardiac examination,
but serves to detect pulse irregularities at an early stage.
) is displayed if you move your body
2) Legs uncrossed
4) Back and arm supported
Picture-06
5.5. Error Indicates
The following symbol will appear on the display when measuring abnormal
SYMBOL
CAUSE
No display
Weak battery or improper
appears
placement
Er 1
Sensor abnormal
Monitor could not detect pulse
Er 2
wave or cannot calculate the
blood pressure data
Er 3
Measurement result is abnormal
Too loose cuff or air leakage
Er 4
Er 5
The air tube is crimped
Er 6
The sensor is sensing great
fluctuation in the pressure
The pressure that the sensor
Er 7
sensing is over the limit
The demarcation is incorrect or the
Er 8
device has not been demarcated
CORRECTION
Replace both batteries with new ones.
Check the battery installation for proper
placement of the battery polarities.
Check if the pump is working or not. If it
is working, then the problem is sensor
abnormal. Please send it to the local
distributor.
start the measurement again.If the error is
still displayed, please send it to local
distributor
Occasionally-measure for one more time/
Always - send it to local distributor
Tie the cuff correctly and make sure the air
plug is properly inserted in the unit
Correct it and make the measurement again
start the measurement again.If the error is
still displayed, please send it to local
distributor
Please send back to the local distributor
Need help?