CONTENTS 1. Introduction and Intended Use....................3 2. Important Information on Blood Pressure and its Measurement...........5 3. Components of your blood pressure monitor.................7 4. Using your Monitor for the First Time..................8 5. Measurement Procedure..................... 9 6. Care and Maintenance......................16 7. Warranty..........................16 8.
1.Introduction and Intended Use The device is intended to measure systolic and diastolic blood pressure as well as the pulse by wrapping around the wrist with cuff circumference ranging from 13.5cm to 19.5cm . The device is used for adult that age is more than 12 years old ,and the intended populations are the patients with hypertension or need blood pressure monitoring.The device is not used for patients under dialysis therapy or on anticoagulant, antiplatelets, or steroids.
Page 5
Warning: • Too frequent measurements can cause injury to the PATIENT due to blood flow interference. • Don’t place the cuff over wound part. • Pressurization of the CUFF can temporarily cause loss of function of simultaneously used monitoring ME EQUIPMENT on the same limb. •...
environment. Caution: To avoid any possibility of accidental strangulation, keep this unit away from children and do not drape tubing around your neck. Caution: To avoid damaging the device, keep this unit away from children and pets. Caution: The standard material used for the bladder and tubing is latex-free. Attention: Self-measurement means control, not diagnosis or treatment.
Page 7
Picture-01 There are six grids in the display of device. Please refer to the picture-01-01. Different grids represent different interval scales of WHO. Blood pressure is very high if your diastolic pressure is above 90 mmHg and/or your systolic blood pressure is over 160 mmHg, while at rest. In this case, please consult your physician immediately.
10. Pulse display 11.WHO symbol 12. Memory symbol 13. Diastolic blood pressure 14.Systolic blood pressure 3.3. Features of Model W03 1. Talking function 2. Double users: 2 x 120 sets memory 3.Date/time display 4. Irregular heartbeat checking 5. Average value function...
7. WHO function 8. Auto power-off 9.Volume adjustment 4. Using your Monitor for the First Time 4.1. System Settings After you load the battery for the monitor,long press the START/STOP button for more than 3s, and then you can start to set. A.
Page 10
factors influence the measurement result. Find time to relax by sitting in an wristchair in a quiet atmosphere for about ten minutes before taking a measurement. • Remove any gwristent that fits closely to your wrist. • Always measure on the same wrist (normally left). •...
Picture-04 5.4. Measure Procedure Refer to picture 05 The monitor is designed to take measurements and store the measurement values in memory for two people using User ID A and User ID B. 1. Sit comfortably in a chair with your feet flat on the floor. 2.
Page 12
To stop the inflation or measurement, push the START/STOP button,the monitor will stop inflating, start deflating, and will turn off After the monitor has detected your blood pressure and pulse rate, the cuff automatically deflates. Your blood pressure and pulse rate are displayed. If getting a suspicious reading after finishing measurement,you should consult the doctor.
Page 13
No display Weak battery or improper Replace both batteries with new ones. Check appears placement the battery installation for proper placement of the battery polarities. Sensor abnormal Check if the pump is working or not. If it is working, then the problem is sensor abnormal. Please send it to the local distributor.
Page 14
5.8 SYMBOL DESCRIPTIONS The following symbols may appear in this manual, on the Blood Pressure Monitor W03, or on it's accessories. Some of the symbols represent standards and compliances associated with the Blood Pressure Monitor W03 and its use. Authorized Representative in the European Community...
Page 15
Follow instructions for use Put up fragile Keep dry Avoid the sun handle gently temperature range 5.9. Memory At the end of a measurement, this monitor automatically stores each result with date and time. Each unit stores 120 sets measurements for 2 users, totally 240 sets(User A and B). Viewing the stored values With the unit off, press the Memory button.
d) To set date and time, follow the procedure described in Section 4.2. Picture-06 Which batteries and which procedure? Use two new, longlife 1.5V AAA batteries. Do not use batteries beyond their expiration date. If the monitor is not going to be used for a prolonged period the batteries should be removed. Using rechargeable batteries You can also operate this instrument using rechargeable batteries.
Device fulfills the stipulations of the International standard IEC60601-1-2 9. Technical Specifications Model: W03 Wight: 101g (batteries and AC adapter is not included) Display: 49.5 x 41.5mm LCD Digital Display Size: 68 (W) x 68 (L) x 30 (H) mm Accessories: 1×Main Device, 1×Cuff, 1×Users manual, 1×Warranty card...
Measuring method: Oscillometric Operating method: Continuous operation Pressure sensor: Resistive Measuring range: DDIA: 15-160mmHg; SYS: 40-240mmHg Pulse: 40 to 199 per minute Cuff pressure display range: 0–295 mmHg Memory: Automatically stores the last 120 measurements for 2 users (total 240) Measuring resolution: 1 mmHg Accuracy: Pressure within ±...
Page 19
RF emission The device is suitable for use in all establishments, Class B including domestic establishments other than domestic CISPR 11 and those directly connected to the public low-voltage Hwristonic emissions power supply network that supplies buildings used for Not applicable domestic purposes.
Page 20
Guidance and manufacture’s declaration – electromagnetic immunity The device is intended for use in the electromagnetic environment specified below. The customer or the user of device should assure that it is used in such an environment. Electromagnetic environment - Immunity test IEC 60601 test level Compliance level guidance...
Page 21
is intended for use in the electromagnetic environment specified below. The customer or the user device of the should assure that it is used in such an environment. device Immunity IEC 60601 test Compliance Electromagnetic environment - guidance test level level Portable and mobile RF communications equipment should be used no closer to any part of the device,...
Page 22
additional measures may be necessary, such as re-orienting or relocating the device. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m. Recommended separation distances between portable and mobile RF communications equipment and the device The device is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled.
The MANUFACTURER should consider reducing the minimum separation distance, based on RISK MANAGEMENT, and using higher IMMUNITY TEST LEVELS that are appropriate for the reduced minimum separation distance. Minimum separation distances for higher IMMUNITY TEST LEVELS shall be calculated using the following equation: Where P is the maximum power in W, d is the minimum separation distance in m, and E is the IMMUNITY TEST LEVEL in V/m.