The printer output will generate the 40 character LCD display test results when the test
is complete. This selection will determine when the 40-character output will print.
When Print Mode = All the printer output will print the test results from every test
performed. When Print Mode = Fail only the test that have failed will be printed.
The Print Number or Serial number will print for every test, followed by the appropriate
test results depending on the Print Mode that has been selected.
Print Number
Please press the SETUP key. This menu selection works in conjunction with Auto
Print. The Print Number screen appears as follows:
Print NO. = 0
Range : 0 - 9999
The print number is used to identify each test result. The number can be set to match
the exact serial number or some portion of actual serial number of the item under test
or just used as an identifier or tracking number. Type in the desired starting number in
the range from 0 -9999. This number will increment as each test is performed. The
next number that will be used can be viewed from this menu at any time after a test has
been completed. This number is not save as part of the non-volatile memory setup
parameters. Therefore each time the power to the instrument is turned off, the
number will be reset to 0.
Form Feed
Please press the SETUP key. The Form Feed screen appears as follows:
Form Feed = O N
<ENTER> to Select
This command forces a form feed after each complete test sequence. When the form
feed is turned off the pages will break whenever the page is determined to be full by the
printer.
Sample Printer Outputs
Sample printer output with the following setup:
Auto Print = ON and
Print Mode = All and
Print No. = 1000
NO. 1000
M 1 Pass
5.00KV
NO. 1001
M 1 Hi-Limit
5.00KV
NO. 1002
70
1.0s
0.000mA
0.5s
10.21mA
Print Mode = OFF