Introduction Congratulations on your purchase of the Extech CG204 Coating Thickness Tester. The CG204 is a portable meter designed for non-invasive coating thickness measurements. The meter uses two measurement methods: magnetic induction (for ferrous metal substrates) and eddy current (for non-ferrous metal substrates).
3. When the meter powers up, the display will prompt for a YES or NO. 4. Press the SET button for YES or the CANCEL button for NO. 5. Note that all data in all memory locations will be erased, in addition to statistics, calibration values, and alarm limits. CG204-EU-EN V2.1 3/11...
Lock out the calibration mode Delete Zero N Clear the zero cal. data (non-ferrous) Delete Zero F Clear the zero cal. data (ferrous) NOTE: Disable the Auto Power OFF feature before lengthy programming to avoid inconvenient automatic power down while programming. CG204-EU-EN V2.1 3/11...
(previously logged readings are not affected); in addition, statistical data will no longer update. If desired, the group data, statistical values, calibration data, and alarm limit values can be deleted using the programming menu. CG204-EU-EN V2.1 3/11...
Page 6
5. User the ARROW BUTTONS to set a High Alarm value 6. Press OK to store the limit and press BACK to return to the menu 7. Perform the same steps for the LOW LIMIT 8. Use the DELETE LIMIT parameter to clear Alarm Limit values CG204-EU-EN V2.1 3/11...
False readings or outliers can then be removed (deleted) using the programming menu. 4. The final reading is derived from a statistical calculation with regard to the meter’s published accuracy specifications. CG204-EU-EN V2.1 3/11...
Ready the supplied metal substrate samples and necessary films (supplied calibration reference films). Set the meter to: a. DIR: (MENU-OPTIONS-Working Mode-Direct) b. CONTINUOUS: (MENU-OPTIONS-Measure Mode-Continuous mode [flashing units indicator] ) c. MAX: (MENU-OPTIONS-LCD Statistic-Maximum). The meter is now ready for calibration. CG204-EU-EN V2.1 3/11...
1 digit change on the display. Multi-point Calibration This method requires performing two or more sequential single point calibrations. For best results, the expected coating thickness measurements should be within the calibration points. CG204-EU-EN V2.1 3/11...
Page 10
Xeff = (Xm - Xo) ±S NOTE: For coatings thicker than 300 μm, the influence of roughness generally is of no importance and therefore it is not necessary to apply the above calibration methods. CG204-EU-EN V2.1 3/11...
The meter’s display will show “FULL” (in the SINGLE measurement mode). In DIRECT mode, when the memory is full, the newest reading will replace the oldest reading and the statistics will be updated. CG204-EU-EN V2.1 3/11...
Err3: Eddy current and Magnetic induction errors Err4, 5, 6: Unused error displays Err7: Thickness error Please contact Extech Instruments if a problem exists. Maintenance Cleaning and Storage Periodically wipe the meter housing with a damp cloth and mild detergent; do not use abra- sives or solvents.