Service Options; Required Skills, Tools, And Components; Reference Documents - Stryker L9000 Service Manual

Light source
Hide thumbs Also See for L9000:
Table of Contents

Advertisement

Introduction
This manual is intended to be used as a service guide for
Stryker repair technicians in the installation, maintenance,
and repair of the Stryker L9000 Light Source. It is meant
to be used in conjunction with the L9000 Light Source
User Manual (1000-401-120) and does not replace existing
documentation.
Service
Options
Factory Service
Due to the complexity of the L9000 Light Source, Stryker
recommends that any malfunctioning system be returned
to Stryker Endoscopy for repair or replacement, where
specialized equipment and technicians are available to
perform repairs while maintaining full product quality and
safety. If service is needed either during or after the warranty
period:
1.
Contact Stryker Endoscopy at 1-800-624-4422,
or contact your local Stryker Endoscopy sales
representative.
2.
Package all the components carefully in the original
shipping container, if possible.
3.
Ship the L9000 Light Source, prepaid and insured, to:
Stryker Endoscopy Customer Service
Attention: Repair Department
5900 Optical Court
San Jose, California 95138
You may request a loaner unit during the repair period.
On-Site Service
Stryker Endoscopy accepts responsibility for the effects on
safety, reliability, and performance of the equipment only if
readjustments, modifications, and repairs have been carried
out exclusively by a person specifically authorized by Stryker
Endoscopy to do so.
On-site repair should be carried out only by qualified
technicians with the proper test equipment listed in this
manual, so that the safety of operators and patients is not
compromised.
In no event shall Stryker Endoscopy be liable for incidental
or consequential damages in connection with or arising from
the performance or use of its products after unauthorized
modification or repair.
Required Skills, Tools, and Components
The repair procedures described in this manual require a
basic set of skills, tools, and replacement components.
Skills
Stryker recommends diagnostic and repair procedures be
performed by authorized, qualified technicians with training
or experience in the following:
«
Basic electronics
technigues
«
Vectorscope operation
+
Bio-Tek Safety Analyzer
«
Multimeter operation
operation
* Oscilloscope operation
Tools
Most of the procedures described in this manual can be
performed using a basic tool kit that includes the items listed
below.
'Basic tools
Advanced tools
»
Phillips screwdriver
»
Multimeter
»
Allen wrench set
¢
Oscilloscope
。 Flathead screwdriver
(20 Mhz or higher)
+
8" adjustable wrench
NTSC/Pal Vectorscope
»
Bio-Tek Model 601
PRO Safety Analyzer
or equivalent current
leakage tester
» Needlenose pliers
» Color video monitor
» Glass fuse puller
Components
Certain replacement components are for in-house use only
and will not be available to any non-Stryker entity.
Before any parts are purchased, an Indemnification Letter
must be signed and submitted to Stryker Endoscopy,
available from Customer Service at 1-800-624-4422.
Stryker reserves the right to incorporate improvements
to the L9000 without notice and will inform customers of
any significant upgrades. All updated parts will be fully
interchangeable with older versions and will offer at least the
same level of quality and performance.
For up-to-date information on upgrades to the product or to
this manual, contact your Stryker representative.
Reference
Documents
None of the in-house Manufacturing Assembly Procedures
(MAPs), Quality Inspection Procedures (QIPs), specialty
tools, jigs, or fixtures listed in this manual are available for
purchase.