  • Page 1 M M X X - - P P R R O O ® ® R R 3 3 O O p p e e r r a a t t i i o o n n s s M M a a n n u u a a l l 6082 6082-209-001 Rev AA.3 2023-03...
  • Page 3 S S y y m m b b o o l l s s Consult instructions for use General warning Caution No pushing Safe working load Catalogue number Serial number M M A A D D E E I I N N U U . . S S . . A A . . Product made in the United States of America Manufacturer Date of manufacture...
  • Page 5 T T a a b b l l e e o o f f C C o o n n t t e e n n t t s s Warning/Caution/Note Definition ..........................2 Summary of safety precautions ..........................2 Introduction ................................4 Product description ..............................4 Indications for use..............................4 Clinical benefits ..............................4...
  • Page 6 • Always install the cot fastener by qualified personnel only. Improper installation could result in injury to the patient or operator. • Always make sure that all cots meet the installation specifications for the Stryker cot fastener system. • Always adjust the rail clamp assembly to match the cot retaining post position for the cot manufacturer and model number.
  • Page 7 • Always use two operators when a cot is occupied. • Always follow proper hand placement on hand grips. Keep all hands clear of the red safety bar pivots when you load or unload the cot or change cot height position. •...
  • Page 8 P P r r o o d d u u c c t t d d e e s s c c r r i i p p t t i i o o n n The Stryker Model 6082 M M X X - - P P R R O O ® R3 is a manual ambulance cot that consists of a platform, including a mattress, mounted on a wheeled X-frame designed to support and transport a maximum weight of 650 lb (294.8 kg) in pre-hospital...
  • Page 9 C C o o n n t t r r a a i i n n d d i i c c a a t t i i o o n n s s None known. E E x x p p e e c c t t e e d d s s e e r r v v i i c c e e l l i i f f e e M M X X P P R R O O R3 has a 5 year expected service life under normal use conditions and with appropriate periodic maintenance.
  • Page 10 Stryker reserves the right to change specifications without notice. The yellow and black color scheme is a proprietary trademark of Stryker Corporation. M M X X - - P P R R O O is designed to conform to the Federal Specification for the Star-of-Life Ambulance (KKK-A-1822-F).
  • Page 11 Competent authority of the European Member State where the user and/or patient is established. To view your operations or maintenance manual online, see https://techweb.stryker.com/. Have the serial number (A) of your Stryker product available when calling Stryker Customer Service or Technical Support. Include the serial number in all written communication.
  • Page 12 S S e e t t u u p p During setup, unpack the cartons and check all items for proper operation. Confirm proper operation before you place the product into service. Remove all the shipping and packaging materials from the product before use. The vehicle patient compartment where the product will be used must have a: •...
  • Page 13 • Always install the cot fastener by qualified personnel only. Improper installation could result in injury to the patient or operator. • Always make sure that all cots meet the installation specifications for the Stryker cot fastener system. • Always adjust the rail clamp assembly to match the cot retaining post position for the cot manufacturer and model number.
  • Page 14 6060-036-018 6060-036-017 6092-036-018 F F i i g g u u r r e e 2 2 – – V V e e h h i i c c l l e e s s a a f f e e t t y y h h o o o o k k t t y y p p e e s s V V e e h h i i c c l l e e c c o o n n f f i i g g u u r r a a t t i i o o n n W W A A R R N N I I N N G G •...
  • Page 15 F F i i g g u u r r e e 3 3 – – D D e e c c k k a a n n d d b b u u m m p p e e r r h h e e i i g g h h t t Deck height (DH) Bumper height Vehicle safety hook...
  • Page 16 6060-036-018 6060-036-017 6092-036-018 F F i i g g u u r r e e 4 4 – – V V e e h h i i c c l l e e s s a a f f e e t t y y h h o o o o k k p p l l a a c c e e m m e e n n t t Sill Floor structure support P P o o s s i i t t i i o o n n i i n n g g o o f f t t h h e e v v e e h h i i c c l l e e s s a a f f e e t t y y h h o o o o k k , , s s i i d d e e t t o o s s i i d d e e...
  • Page 17 I I n n s s t t a a l l l l i i n n g g t t h h e e v v e e h h i i c c l l e e s s a a f f e e t t y y h h o o o o k k Before vehicle safety hook installation, the certified mechanic should plan for the placement of the vehicle safety hook in the rear of the vehicle patient compartment.
  • Page 18 F F i i g g u u r r e e 5 5 – – S S a a f f e e t t y y b b a a r r s s e e c c u u r r e e d d i i n n t t h h e e v v e e h h i i c c l l e e s s a a f f e e t t y y h h o o o o k k F F i i g g u u r r e e 6 6 –...
  • Page 19 C C o o t t p p o o s s i i t t i i o o n n s s Position 1 - Use for patient transfer Position 2 - Use for patient transfer/cot rolling Position 3 - Use for patient transfer/cot rolling Position 4 - Use for patient transfer/cot rolling Position 5 - Use for patient transfer/cot rolling Position 6 - Use for patient transfer/cot rolling...
  • Page 20 Position 7 - Use for patient transfer/cot rolling W W A A R R N N I I N N G G - - Do not operate the product with the breakaway head section lowered to avoid injury to the patient or operator or damage to the product.
  • Page 21 Always load or unload an occupied cot with a minimum of two trained operators. Two operators must be present when a cot is occupied. Stryker recommends that both operators are at the foot end to reduce the load on each operator. One or two operators can lift from the foot end of the cot.
  • Page 22 To release a wheel lock, press down on the top of the pedal with your foot or lift up on the pedal with your toe. The top of the pedal will rest against the caster frame when you release the wheel lock. N N o o t t e e - - Wheel locks help prevent the product from rolling while unattended.
  • Page 23 F F i i g g u u r r e e 9 9 – – F F o o o o t t s s e e c c t t i i o o n n r r e e s s t t r r a a i i n n t t s s F F i i g g u u r r e e 8 8 –...
  • Page 24 F F i i g g u u r r e e 1 1 2 2 – – S S h h o o r r t t e e n n t t h h e e r r e e s s t t r r a a i i n n t t N N o o t t e e •...
  • Page 25 L L o o a a d d i i n n g g a a c c o o t t i i n n t t o o a a v v e e h h i i c c l l e e w w i i t t h h a a n n a a n n t t l l e e r r s s t t y y l l e e c c o o t t f f a a s s t t e e n n e e r r Always load an occupied cot with a minimum of two trained operators.
  • Page 26 U U n n l l o o a a d d i i n n g g a a c c o o t t f f r r o o m m a a v v e e h h i i c c l l e e w w i i t t h h a a n n a a n n t t l l e e r r s s t t y y l l e e c c o o t t f f a a s s t t e e n n e e r r Always unload an occupied cot with a minimum of two trained operators.
  • Page 27 P P o o s s i i t t i i o o n n i i n n g g o o p p e e r r a a t t o o r r s s a a n n d d h h e e l l p p e e r r s s If equipped with the right hand release option: C C h h a a n n g g i i n n g g l l e e v v e e l l s s R R o o l l l l i i n n g g...
  • Page 28 R R a a i i s s i i n n g g o o r r l l o o w w e e r r i i n n g g t t h h e e c c o o t t w w i i t t h h t t w w o o o o p p e e r r a a t t o o r r s s Always raise or lower an occupied cot with a minimum of two trained operators.
  • Page 29 R R a a i i s s i i n n g g o o r r l l o o w w e e r r i i n n g g t t h h e e c c o o t t w w i i t t h h o o n n e e o o p p e e r r a a t t o o r r You can raise or lower an unoccupied cot into a vehicle with one operator.
  • Page 30 R R a a i i s s i i n n g g o o r r l l o o w w e e r r i i n n g g t t h h e e b b r r e e a a k k a a w w a a y y h h e e a a d d s s e e c c t t i i o o n n Fold the breakaway head section down to shorten the length of the cot and allow for improved mobility and maneuvering in limited spaces (elevators, halls, etc.).
  • Page 31 To attach the defibrillator platform: 1. Place the defibrillator platform in the stored position (Figure 16). 2. Open and expand the defibrillator platform legs (Figure 17). F F i i g g u u r r e e 1 1 6 6 – – S S t t o o r r e e d d p p o o s s i i t t i i o o n n F F i i g g u u r r e e 1 1 7 7 –...
  • Page 32 8. Place the defibrillator on the defibrillator platform. 9. Secure the straps (E) to hold the defibrillator on the defibrillator platform (Figure 20). F F i i g g u u r r e e 2 2 0 0 – – A A t t t t a a c c h h d d e e f f i i b b r r i i l l l l a a t t o o r r p p l l a a t t f f o o r r m m t t o o t t h h e e c c o o t t N N o o t t e e - - If the defibrillator platform is not attached to the product after you connect both latch hooks, unlatch the tab.
  • Page 33 F F i i g g u u r r e e 2 2 2 2 – – T T w w o o - - s s t t a a g g e e I I V V p p o o l l e e s s t t o o r r a a g g e e p p o o s s i i t t i i o o n n F F i i g g u u r r e e 2 2 1 1 –...
  • Page 34 F F i i g g u u r r e e 2 2 4 4 – – T T h h r r e e e e - - s s t t a a g g e e I I V V p p o o l l e e s s t t o o r r a a g g e e p p o o s s i i t t i i o o n n F F i i g g u u r r e e 2 2 3 3 –...
  • Page 35 A A c c c c e e s s s s o o r r i i e e s s a a n n d d p p a a r r t t s s These accessories may be available for use with your product. Confirm availability for your configuration or region. Call Stryker Customer Service: 1-800-327-0770. N N a a m m e e...
  • Page 36 C C l l e e a a n n i i n n g g a a n n d d d d i i s s i i n n f f e e c c t t i i n n g g w w i i t t h h S S i i d d e e K K i i c c k k ® ® For United States only. Confirm availability for your configuration or region. Call Stryker Customer Service: 1-800-327- 0770. Stryker’s preferred wipes (2060-000-001 6'' x 10'' or 2060-000-002 9'' x 12'') and spray (2070-000-001) include the following active ingredients: •...
  • Page 37 C C l l e e a a n n i i n n g g W W A A R R N N I I N N G G - - Always use any appropriate personal protective equipment while power washing to avoid inhaling contagion. Power washing equipment may aerate contamination.
  • Page 38 P P r r e e v v e e n n t t i i v v e e m m a a i i n n t t e e n n a a n n c c e e O O p p e e r r a a t t i i o o n n S S c c h h e e d d u u l l e e P P r r o o c c e e d d u u r r e e...
  • Page 39 F F i i g g u u r r e e 2 2 5 5 – – B B a a s s e e l l u u b b r r i i c c a a t t i i o o n n l l o o c c a a t t i i o o n n s s 4.
  • Page 40 Stryker Medical 3800 E. Centre Avenue Portage, MI 49002 6082-209-001 Rev AA.3 2023-03 WCR: AA.35...