6. PREPARATION FOR USE

INSPECTION
Prior to installing the Q55 Ultrasonic Homogenizer, perform a visual inspection to de-
tect any evidence of damage, which might have occurred during shipment. Before dis-
posing of any packaging material, check it carefully for small items.
The Q55 Ultrasonic Homogenizer was carefully packed and thoroughly inspected be-
fore leaving our factory. The carrier, upon acceptance of the shipment, assumed respon-
sibility for its safe delivery. Claims for loss or damage sustained in transit must be sub-
mitted to the carrier.
If damage has occurred, contact your carrier within 48 hours of the delivery date. DO
NOT OPERATE DAMAGED EQUIPMENT. Retain all packing materials for future shipment.
ELECTRICAL REQUIREMENTS
The Q55 Ultrasonic Homogenizer requires 24 volts DC. A universal 98-264 V @ 47~63Hz
AC, to DC switch-mode power supply capable of delivering 24 volts at 3.75 amperes is
supplied with your system.
If it is necessary to convert the unit for different voltage operation, proceed as follows:
1.
Ensure that the power cord is not connected to the electrical outlet.
2.
Change the electrical power cord or add a properly grounded universal AC socket
adapter as required.
For your personal safety, do not, under any circumstances, defeat the
grounding feature of the power cord by removing the grounding prong.
INSTALLING THE Q55 ULTRASONIC HOMOGENIZER
The Q55 Ultrasonic Homogenizer should be installed in an area that is free from exces-
sive dust, dirt, explosive and corrosive fumes, and extremes of temperature and humid-
ity. If processing flammable liquids, use an approved fume hood and do not place the
power supply in the fume hood.
When positioning the unit, be sure to leave adequate space behind the unit so that all
connections can be easily disconnected.
WARNING