System Inspection; Leak-Testing The Connections - MAXIMATOR SLGP 3 Series Installation And Operating Manual

Air driven liquefied gas pump
Table of Contents

Advertisement

Maintenance
7.2.1

System inspection

The following section explains how to check the pump for proper function:
Qualifications
Type of mainte-
nance
Interval
PPE
1.
2.
3.
4.
7.2.2

Leak-testing the connections

The following section explains how to check the connections for leaks:
Qualifications
Type of mainte-
nance
Interval
31
Description
Operating the system
Check
weekly
Safety goggles
Hearing protection
Shut off the fluid outlet and adjust p
tem.
The pump stops automatically when the final pressure is reached (holding
time 60 s).
Vent p
.
L
p
does not drop by more than 10% (holding time 30 s).
B
Set p
to approx. 50 % of the value from the first step and slowly relieve p
L
The pump starts up automatically.
If the inspection does not reveal any abnormalities, it will be safe to contin-
ue using the pump.
In case of abnormalities, consult with the maintenance staff.