Calibration motor control——control the x, y calibrate motor.
Calibration Offset——show the angle offset after calibration
Calibration Method——change the calibration of multi-fiber

4.4 Software operation instruction

4.4.1 Single fiber calibration

Click calibrate on single measure interface to enter single fiber calibration interface
Finish the calibration following the instruction of software.
Awareness in calibration:
1. Keep tool clean
2. Rotate the tool 60 degree each time, too large or too small will cause the calibration fail.
3. Keep focused in each measurement
4. Measure the reliability after calibration. Measure the result on rotation and make sure the
apex offset difference is lower than 3um.

4.4.2 Multi-Fiber Calibration: Two step

The default calibration method is two steps method.
Click Calibrate to enter the calibration interface.
Figure 4-15Single calibration interface
25
MTPro operation manual