Instructions for use KAVOLUX 1410 C Contents Contents User instructions ............................3 User guide ............................3 1.1.1 Abbreviations ........................3 1.1.2 Symbols ..........................3 1.1.3 Target group ........................3 Service ............................3 Warranty terms and conditions ......................4 Transportation and storage ......................4 1.4.1 Currently valid packaging regulations ................
Page 4
Instructions for use KAVOLUX 1410 C Contents Cleaning and disinfecting the handles ................... 32 Sterilisation ............................ 33 7.4.1 Sterilising the handles ...................... 33 Safety checks ............................34 Repair ..............................35 Exchanging the dental reflector ..................... 35 Exchange the halogen lamp ......................36 Exchange the gas spring .......................
This document is for dentists, the dental office personnel, and the service techni‐ cian. 1.2 Service Technical Service +49 (0) 7351 56-2500 Service.Einrichtungen@kavo.com Please refer to the serial number of the product in all requests! For further information, please visit: www.kavo.com 3/44...
Claims from this warranty can only be asserted when the transfer form (copy) be‐ longing to the product has been sent to KaVo, and the original can be presented by the operator or user. 1.4 Transportation and storage 1.4.1 Currently valid packaging regulations...
4. Report the damage to the shipping company. 5. Report the damage to KaVo. 6. You must contact KaVo before returning a damaged product. 7. Send the signed delivery receipt to KaVo. If the product is damaged but there was no discernable damage to the packaging upon delivery, proceed as follows: 1.
Instructions for use KAVOLUX 1410 C 1 User instructions | 1.4 Transportation and storage Note Failure on the part of the recipient to comply with one of the above obligations will mean that the damage will be considered to have arisen following delivery (in accordance with CMR law, Chapter 5, Art.
2.2 Purpose – Proper use This KaVo product is an operating light for illuminating the oral cavity during den‐ tal treatment. Install the KAVOLUX 1410 C directly on the ESTETICA E50, E70 or E80 treat‐...
The overarching guidelines and/or national laws, national regulations and the rules of technology applicable to the startup and use of the KaVo product for the intended purpose are to be applied and complied with.
This must be done in strict compliance with all applicable national regulations. Questions on proper disposal of the KaVo product can be answered by the KaVo branch. 2.2.3 Disposal of electronic and electrical devices Note According to EC directive 2002/96 concerning used electrical and electronic de‐...
3 Product description | 3.1 Treatment light 3 Product description Note The KAVOLUX 1410 C operating light is only suitable for installation on the KaVo treatment centres, ESTETICA E80, E70, and E50. 3.1 Treatment light Components of the KAVOLUX 1410 C: ①...
Instructions for use KAVOLUX 1410 C 3 Product description | 3.2 Dimensions and swing range 3.2 Dimensions and swing range CAUTION Collision with persons or pieces of equipment. Collisions could be caused by the required degrees of freedom and the large swivel range ▶...
Instructions for use KAVOLUX 1410 C 3 Product description | 3.3 Rating plates Swing ranges (in mm) 3.3 Rating plates The rating plate of the operating light is located on the top side of the swivel arm. RE F: Rating plate of the KAVOLUX 1410 C-L 12/44...
Instructions for use KAVOLUX 1410 C 3 Product description | 3.4 Technical data and requirements RE F: Rating plate of the KAVOLUX 1410 C-R ① Device type ⑥ Material number ② Serial number ⑦ CE mark ③ Material number ⑧ Disposal instructions (see "Proper use") ④...
Instructions for use KAVOLUX 1410 C 4 Assembly | 4.1 Packaging 4 Assembly 4.1 Packaging 4.1.1 Transportation and storage The packaging consists of styrofoam and is itself packaged in Superwell card‐ board. The instructions printed on the outside are for transportation and storage, and have the following meaning: 1.
Instructions for use KAVOLUX 1410 C 4 Assembly | 4.2 Mounting the device ▶ Remove the styrofoam filling ④ from the operating light. 4.2 Mounting the device 4.2.1 General instructions on mounting the unit Note Comply with the provisions of IEC 601. Turn off the main switch before mounting.
Instructions for use KAVOLUX 1410 C 4 Assembly | 4.2 Mounting the device 4.2.2 ESTETICA E50 Install the mounting pole 16/44...
Page 19
Instructions for use KAVOLUX 1410 C 4 Assembly | 4.2 Mounting the device ▶ Removing the rear service flap ▶ Removing the cover of the connection point. ▶ Screw off the cover ①. ▶ Loosen screw ② below the main switch. ▶...
Page 20
Instructions for use KAVOLUX 1410 C 4 Assembly | 4.2 Mounting the device ▶ Loosen screws ③ on the disinfection container and remove the container ④. ▶ Ensure that the protective earth screw ⑥ and fastening screws ⑤ and ⑦ do not reach into the light holder.
Page 21
Instructions for use KAVOLUX 1410 C 4 Assembly | 4.2 Mounting the device ▶ Align the body of the unit so that the mounting pole is exactly vertical (see de‐ vice installation instructions). Check with a spirit level. ▶ Push down the cover ring on the mounting pole. Install the swivel arm ▶...
Page 22
Instructions for use KAVOLUX 1410 C 4 Assembly | 4.2 Mounting the device ▶ Guide the lines through the hole in the mounting pole and then downward until they exit the bottom end of the mounting pole. ▶ Place the cap (Mat. no. 1.003.7006) on the swinging arm. 20/44...
Page 23
Instructions for use KAVOLUX 1410 C 4 Assembly | 4.2 Mounting the device Connecting the lines ▶ Plug-on the voltage supply (24 V AC) of the operating light at slot X13 (0 V) ② and X13 (24 V) ③. CAN-R Mat.
Instructions for use KAVOLUX 1410 C 4 Assembly | 4.2 Mounting the device 4.2.3 ESTETICA E70 / ESTETICA E80 Install the mounting pole 22/44...
Page 25
Instructions for use KAVOLUX 1410 C 4 Assembly | 4.2 Mounting the device ▶ Remove device cladding (see assembly instructions of the treatment unit). ▶ Slide the cover ring ① onto the mounting pole. ▶ Insert the mounting pole into the light holder as far as it will go ②. ▶...
Instructions for use KAVOLUX 1410 C 4 Assembly | 4.3 Set the lamp voltage Connect the lines ▶ Screw the earth protector to the earth terminal. ▶ Run the lines for the 24 V supply and control in the device body. ▶...
▶ After setting the correction value check the voltage at the light. 4.3.2 ESTETICA E70 to Ser. No.: 20100000 Note The KaVo Service Tester (Mat. no. 1.003.2212) is required to make adjustments in service mode. ▪ Service tester: Service tester I²C Mat. no. 1.003.2212...
Instructions for use KAVOLUX 1410 C 4 Assembly | 4.3 Set the lamp voltage ▶ To start service mode, press and hold the "Decrease value" and "Increase val‐ ue" keys. ▶ Press the foot pedal. Service mode U1 is shown on the display. ▶...
Page 29
Instructions for use KAVOLUX 1410 C 4 Assembly | 4.3 Set the lamp voltage ▶ Press the "Instrument menus" button until "U09" is shown on the display. Note The service mode U09 "Treatment light correction values" can only be imple‐ mented as of E-PROM version 2.1.0 on the MEMOdent.
Instructions for use KAVOLUX 1410 C 5 Mechanical settings | 5.1 Adjusting the swinging arm brake 5 Mechanical settings 5.1 Adjusting the swinging arm brake Note The swinging arm brake determines the ease with which the swinging arm ro‐ tates. ▶...
Page 31
Instructions for use KAVOLUX 1410 C 5 Mechanical settings | 5.2 Adjusting the light head brake Note To remove the protective shield, do not grab the windows; instead, grasp the plastic frame. Do not touch the dental reflector with your fingers. ▶...
Instructions for use KAVOLUX 1410 C 6 Operation 6 Operation Note The operation of the KAVOLUX 1410 C operating light is described in the In‐ structions for Use of the treatment unit. 30/44...
Instructions for use KAVOLUX 1410 C 7 Rehabilitation methods according to DIN EN ISO 17664 | 7.1 General care instructions 7 Rehabilitation methods according to DIN EN ISO 17664 7.1 General care instructions CAUTION Damage to product as a result of improper cleaning. Malfunctions, unclean surfaces and risk of infection for operator and patients.
Instructions for use KAVOLUX 1410 C 7 Rehabilitation methods according to DIN EN ISO 17664 | 7.3 Cleaning and disinfecting the handles ▶ Clean the lamp, reflector, and protective shield with lukewarm water and a soft cloth free of lint. ▶...
Instructions for use KAVOLUX 1410 C 7 Rehabilitation methods according to DIN EN ISO 17664 | 7.4 Sterilisation ▶ Disinfect the white handles (Mat. no. 1.001.4472) and the grey handles (Mat. no. 1.002.4489) with a disinfectant wipe or thermodisinfection (slight dis‐ colouration possible).
Instructions for use KAVOLUX 1410 C 8 Safety checks 8 Safety checks The following safety checks must be done every 24 months: ▪ Check the cover/chip protection of the halogen lamp. ▪ Check the fastening of the reflectors ▪ Check the support arm system, the joints, the gas spring and handles for me‐ chanical strength and damage.
Instructions for use KAVOLUX 1410 C 9 Repair | 9.1 Exchanging the dental reflector 9 Repair 9.1 Exchanging the dental reflector Note The dental reflector consists of two halves so that perhaps only one-half needs to be exchanged. Plastic dental reflector Mat. no. 0.740.4772. Note To remove the protective shield, do not grab the windows;...
Instructions for use KAVOLUX 1410 C 9 Repair | 9.2 Exchange the halogen lamp CAUTION The reflector shells can break if they are too tightly clamped. Potential consequences ▶ Lightly clamp the reflector shells with the outer setscrews and then fix them with the middle setscrews.
Page 39
Instructions for use KAVOLUX 1410 C 9 Repair | 9.2 Exchange the halogen lamp ▶ Remove the halogen lamp from the guide. Note Do not touch the glass bulb of the new halogen lamp with your fingers. ▶ Screw the wires of the new halogen lamp tightly to the luster terminal. ▶...
Page 40
Instructions for use KAVOLUX 1410 C 9 Repair | 9.2 Exchange the halogen lamp ▶ Insert the new halogen lamp into the guide. ▶ Let the retention spring lock in place. ▶ Mount the protective cover. 38/44...
Instructions for use KAVOLUX 1410 C 9 Repair | 9.3 Exchange the gas spring 9.3 Exchange the gas spring Note Only trained technicians may change the spring. 39/44...
Instructions for use KAVOLUX 1410 C 10 Eliminating disturbances 10 Eliminating disturbances Note If disturbances cannot be located or eliminated by using this troubleshooting method, a technician authorised by KaVo must try and eliminate the problem. Malfunction Cause Remedy Treatment unit and oper‐...
Instructions for use KAVOLUX 1410 C 11 Specifications on the electromagnetic compatibility in accordance with EN 60601-1-2 | 11.1 Guidelines and manufacturer's statement - Electromagnetic uses 11 Specifications on the electromagnetic compatibility in accord‐ ance with EN 60601-1-2 11.1 Guidelines and manufacturer's statement - Electromagnetic uses The operating light is designed for operation in the environment defined below.
Instructions for use KAVOLUX 1410 C 11 Specifications on the electromagnetic compatibility in accordance with EN 60601-1-2 | 11.3 Guidelines and manufacturer's statement - Electromagnetic immunity Interference immunity IEC 60601 test levels Compliance level Electromagnetic environ‐ tests ment - Guidelines Electrostatic discharge ±...
Page 45
Instructions for use KAVOLUX 1410 C 11 Specifications on the electromagnetic compatibility in accordance with EN 60601-1-2 | 11.3 Guidelines and manufacturer's statement - Electromagnetic immunity Interference resistance IEC 60601 test levels Compliance Electromagnetic environment - tests level Guidelines Conducted HF distur‐ 150 kHz to 80 Portable and mobile radio sets bances according to...
Instructions for use KAVOLUX 1410 C 11 Specifications on the electromagnetic compatibility in accordance with EN 60601-1-2 | 11.4 Recommen‐ ded safe distance between portable and mobile HF telecommunications equipment and the KaVolux 1410 performance characteristics are observed, for example, a change in alignment or a different location of the .
Need help?
Do you have a question about the LUX 1410 C and is the answer not in the manual?