MODEL 6oOA ELXC!CBOMEZER
CIRCUIT DISCUSSION
drop divided by the current measured.
It will
be rarely
necessary
to correct for the input drop of the electrometer
due to the excel-
lent voltaRe sensitivity
of the Model
600~.
It is advisable to use the NOW micro-microammeter connection
for the measurement of leakage resistance
of capacitors
in this
manner, since instability
is likely
to occur using the FAST con-
nection.
However, in cases where the capacity shunted across the
sample is small, it will
be possible to realize
a considerable
increase in speed of response by utilizing
the FAST connection.
Some precautions
are recommended when testing
capacitors.
Be sure
that capacitors
have discharged before removing from test circuit.
With the ZERO CHECK switch depressed,the input is shorted to
ground through 1 megohm, providing
a discharge path for the cana-
citor.
The reading: time may be shortened if the capacitor
is al-
lowed to charge through a low impedance position
on the AMPERES
switch before the appropriate
measuring resistor
is inserted
in
the circuit.
It should be further
noted that capacitor measurement is likely
to be a slow process in any case due to the fact that it may take
considerable
time for the molecular orientation
of the dielectric
to take place at the testing
potential.
It may take minutes or
even hours in some cases to achieve a stable reading.
4-5