Maintenance and Troubleshooting manual
00809-0200-3104
7. Carefully install the transducer holder assembly into the transducer mount. Ensure
8. Install the keyed transformer module into the transducer holder (see
Rosemount 3415, 3416 and 3417
the threads of the holder and mount are correctly aligned. Use a 1¼-in. socket
and screw the transducer assembly into the mount. Tighten to securely seat the
assembly in the mount.
Do not overtighten (see
Figure
a) Apply a small amount of Molykote 111 to the transformer module O-ring.
b) Insert the keyed transformer module into the back end of the transducer
holder.
Figure 3-3: T-22 transducer assembly, holder, transformer assembly, retainer,
cable nut, and chordset
A. T-21 transformer assembly (W-01 PN 1-360-03-090) or T-22 transformer assembly
(W-02 PN 1-360-03-110)
B. Transducer holder (type - H1 PN 1-360-01-128, H2 PN 1-360-01-228)
C. Mount (Inconel mount and holder)
D. Transducer port (meter body)
E. Transducer retainer (standard PN 1-360-01-958) (high temperature PN
1-360-01-978)
F. Cable nut and chordset:
•
212 ° F (100 °C) -5-ft. (1.5 m) length (PN 3-3400-190)
•
212 ° F (100 °C) -15-ft. (4.6 m) length (P/N 3-3400-194)
3-3).
Meter repairs
December 2023
Figure
3-3).
41