6.2
Technical Specifications
Stimulator Output
Output Range and Accuracy:
Stimulation Frequency:
Pulse Characteristics:
Inter-pulse Spacing:
0%-100% of Maximum Voltage± 1%
0.25Hz in Bistim Mode
Magstim Bistim
2
Operating Manual
0.50Hz at 01% to 49% Power Output in simultaneous discharge mode
0.33Hz at 50% to 75% Power Output in simultaneous discharge mode
0.25Hz at 76% to 100% Power Output in simultaneous discharge mode
1 00µs rise time, 1 ms duration
0 - 999ms or 0- 99.9ms
Note: Magnetic field strength and maximum number of stimulations produced before coil exceeds
operating temperature are dependent on the stimulation coil being used. Please refer to the
appropriate coil operating manual.
Capacitor Life Expectancy
Life expectancy:
Isolated Interface
Logic High Voltage:
Max Voltage:
Logic Low Voltage:
Min Voltage:
2 x 10
6
discharges at 70% power level
8 x 10
5
discharges at 80% power level
4 x 10
5
discharges at 90% power level
2 x 10
5
discharges at 100% power level
>4.0V
5.3V
<0.8V
-0.3V
All voltages are with respect to GND.
Power Supply
Power:
Fuse(s):
General Specifications
Master and Slave Units:
Dimensions:
Weight:
Stacking limit:
Bistim
2
Connecting Module:
Dimensions:
Weight:
Software
Controller Board Software:
UI Controller Software:
110V-120V- 50-60Hz;
220V-240V- 50-60Hz;
100VA during standby
2000VA peak and 600VA average during discharge
2 x T1 0A 20x5mm for 11 0V-120V-
2 x T5A 20x5mm for 220V-240V-
460mm x 160mm x 375mm
20.4kg
4 Units stacked on top of each other
135mmx x 81mm x 165mm
1.62kg
Version 1.9
(Software ref: 3056)
Version 1.1
(Software ref: 3026)
Please note: All specifications are subject to alteration.