0
3.1.11
Output Control (12)
Magstim Bistim
2
Operating Manual
The output control allows the power level displayed on the power output display to be adjusted.
Rotating the control clockwise increases the power level. Rotating anti-clockwise decreases the
power level. Each complete 360° rotation changes the output level by approximately 20%.
3.1.12
TriggerButton(13)
The Magstim Bistim
2
can be triggered to deliver the stimulus by pressing the yellow trigger button
along with one of the trigger buttons located on the coil. The instrument will continuously charge and
discharge if the trigger buttons are pressed and held.
3.1.13
System Stop Button (14)
The Magstim Bistim
2
can be disarmed by pushing the red stop button. In this mode the disarmed
indicator will illuminate and the instrument will discharge internally.
Note: The Magstim Bistim
2
will disarm if the coil is disconnected, becomes faulty, the surface
temperature of the connected coil has risen above its set limit, the system stop button is activated, the
instrument detects an internal fault, or if the ready symbol has been illuminated for over 1 minute
without the instrument having been triggered.
3.1.14
Coil Output Socket (15)
This socket is used to connect the stimulating coil to the Magstim Bistim
2
system.
Note: It is normal for the pin indicated on the label to protrude from the coil socket on the master and
slave stimulator units. This is an intentional design feature.
Figure 3.2: Coil output socket front view
Figure 3.3: Coil output socket angled view