Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The Magstim Bistim
2
is intended for use in the electromagnetic environment specified below. The customer or the user of the equipment
should assure that it is used in such an environment.
Immunity Test
IEC 60601
Compliance
Electromagnetic environment - guidance
Test Level
Level
Electrostatic
± 6 kV contact
Meets requirement
Floors should be wood, concrete or ceramic tile. If floors are
discharge {ESD)
± 8 kV air
covered with synthetic material, the relative humidity should be at
least 30%.
IEC 61000-4-4
Electrical fast
± 2 kV for power
Meets requirement
Mains power quality should be that of a typical commercial or
Transient/ burst
supply lines
hospital environment.
IEC 61000-4-11
± 1 kV for input/
output lines
Surge
± 1 kV differential
Meets requirement
Mains power quality should be that of a typical commercial or
mode
hospital environment.
IEC 61000-4-5
± 2 kV common mode
Voltage dips, short
<5%
Ur
Meets requirement
Mains power quality should be that of a typical commercial or
interruptions and
(>95 % dip in Ur) for
hospital environment. If the user of the Magstim Bistim
2
requires
voltage variations
0,5 cycle
continued operation during power interruptions, it is
on power supply
recommended that the Magstim Bistim
2
be powered from an
input lines.
40 %
Ur
interruptible power supply.
(60 % dip in Ur) for 5
IEC 61000-4-11
cycles
70 %
Ur
(30 % dip in Ur) for 25
cycles
<5%
Ur
(>95 % dip in Ur) for 5
sec
Power Frequency
3A/m 50Hz
Meets requirement
Power frequency magnetic fields should be at levels
(50/60Hz)
characteristic of a typical location in a typical commercial or
Magnetic Field
hospital environment
!EC 61000-4-8
Radiated RF
80MHz - 2.5GHz
Meets requirement
Equipment should only be used in the vicinity of other equipment
Immunity
2Hz 80% amplitude
compliant with EN60601-1-2.
EN61000-4-3
modulation
Conducted RF
0.15MHz - 80MHz
Meets requirement
Equipment should only be used in the vicinity of other equipment
Immunity
2Hz 80% amplitude
compliant with EN60601-1-2.
EN61000-4-6
modulation
NOTE
U. is the a.c. mains voltage prior to application of the test level
© The Magstim Company Limited