0
0
3.1.4
Output Display (4)
Magstim Bistim
2
Operating Manual
The output display consists of 3 lines. The first and second lines display the power output as a
percentage of maximum of the master and slave units respectively. The third line displays the inter-
pulse spacing. The default value displayed after initially switching on the master and slave units are
30% for both power outputs and 10ms for the inter-pulse spacing.
If an error code needs to be displayed, the power output is replaced by error code(s). These are
detailed in Appendix A - System Error Codes.
3.1.5
Coil Indicator (5)
This green LED will illuminate if there is a coil detected as being connected correctly to the Magstim
Bistim
2
3.1.6
Coil Temperature Indicators (6)
These LED's represent the surface temperature of the coil. The LED's illuminate sequentially as the
coil temperature rises, beginning with the bottom green LED and ending with the amber LED. When
the amber LED is lit, it means that the coil is getting close to its maximum operating temperature. As
soon as it reaches its maximum operating temperature, the replace coil indicator will illuminate and
the Magstim Bistim
2
will disarm itself.
NB If the temperature of the coil is below 5°C, the system will not operate. The Magstim Bistim
2
will
remain in its inactive default condition until it detects that the coil temperature has risen above 5°C
3.1.7
Replace Coil Indicator (7)
This red LED will illuminate if any of the following conditions are present:
1.
The coil connected is not compatible with the Magstim Bistim
2
2.
The surface temperature of the connected coil has risen above its set limit.
3.
The connected coil is faulty.
4.
There is no coil connected.
If this LED is illuminated the Magstim Bistim
2
will not arm until the reason for the indicator being lit is
resolved.
3.1.8
Disarmed Indicator (8)
This LED is illuminated when the Magstim Bistim
2
has been put into its safe disarmed state.
3.1.9
Ready Indicator (10)
When the Magstim Bistim
2
is armed, this LED will illuminated when the Magstim Bistim
2
is charged
and ready to be triggered.
3.1.10
Run Button (11)
The Magstim Bistim
2
can be armed pushing the green run button. The Magstim Bistim
2
will charge to
the power level indicated on the output display. The Magstim Bistim
2
will not arm if the replace coil
indicator is illuminated.
© The Magstim Company Limited