MAINTENANCE
MODELS 610B,
610BR ELECTROMETERS
4.
Check the next
stage
by connecting
the voltmeter
across
the
collectors
of 9103 end Q104 (Figure
21),
end adjusting
the COARSE and MEDIUM ZERO Controls
for
null.
If
a null
is
not
reached,
check
this
stage
and for
shorts
in
the
base circuit
of Q105 and QlO6.
5.
Connect
the voltmeter
across
the
collectors
of Q105 and Ql06.
Adjust
the
FINE ZERO Control
for
null.
If
null
is
reached,
the
dc amplifier
is
operating
correctly
and the
trouble
is
in the
output
stage
or the
feedback
stage.
6.
Check the
output
stage
cathode
resistor,
Rl8S (Figure
24))
end capa-
citor,
Cl13
(Figure
23),
for
a possible
short.
7.
The feedback
Loop includes
the multiplier
resistors,
RlS6 through
Rl66,
the
recorder
output
resistors,
R180 on 3V position
or
R174 and R179 on 1 MA
position,
and the meter.
An opening
of any of
these
components
prevents
zero-
ing
of the meter.
An open multiplier
resistor,
however,
prevents
zeroing
for
only
that
particular
multiplier
setting.
4-s.
234-VbLT
OPEBATION.
The Model
610B is
shipped
for
use with
a 117-volt
power
source
unless
otherwise
ordered.
To convert
it
for
234-volt
sources,
use a screwdriver
to change
the
slide
switch
on the beck
panel
to 234.
Change
the
fuse
from
l/4
ampere
to l/8
ampere.
No other
adjustment
is necessary.
To switch
from
234 to 117-volt
operation,
reverse
the
procedures.
4-6.
CALIBRATION
PROCEDURES.
'a.
The following
procedures
are
recoranended
for
calibrating
and adjusting
the mdel
610B.
Use the
equipment
reconrmended
in Table
3.
If
proper
facilities
are not
avaflable
or if
difficulty
is
encountered,
contact
Keithley
Instruments,
Inc.,
or its
representatives
to arrange
for
factory
calibration.
b.
Procedures
are
covered
for
the
following:
grid
current
check,
dc ampli-
fier
balance
adjustment,
high-megohm
resistor
verification,
calibration
of
the
ohms ranges,
meter
zero
calibration
and
accuracy
check.
1.
Check grid
current
(paragraph
4-7)
et
regular
intervals
to make sure
the
electrometer
tubes
are
functioning
correctly.
Instrument
Use
Keithley
Instruments
Model
241
Calibrate
meter
zero
and verify
Regulated
High
Voltage
Supply
voltage
range
accuracy.
Keithley
Instruments
Model
261 Picoamp-
Calibrate
current,
resistance
gre
spurce*
fr#m
LO-L4 to LOe4 ampere,
and charge
ranges.
0
1.5/,.
Keithley
Instruments
Model
370 Recorder
Zero
drift
check.
Keithley
Instruments
Model
515
Verify
high-megohm
resistors
in
Megohm Bridge
Range Switch.
KeTCh1eyRrsrrnment.s
Model
662
Calibrate
resisrmrre-ranges
and meter
zero.
TABLE 3.
Equipment
Recommended for
Model
610B Calibration.
Use these
inetru-
ments
or their
equivalents.
32
0667R