MODELS bLOB, 61OBR ELECTROMETERS
OPERATION
2-7.
Resistance
Measurements.
a.
The Model
610B can measure
resistances
by three
methods.
1.
In the normal
or two-terminal
method
(annneter-voltmeter)
, the
ELec-
trometer
measures
the voltage
drop
across
the unknown
sample
as a known,
constant
current
flows
through
it.
The voltage
drop
is
proportional
to
the
resistance
of the
sample.
to LO11 ohm ranges.
This
method
is
the
simplest
for
the
100
2.
Above
LOLL ohms or to prevent
leakage,
the guarded
method
is better.
It
results
in faster
response
speeds
and also
nullifies
Leakage
errors
across
the
Electrometer
input,
since
the potential
across
the
input
ter-
minal
is
small.
3.
In
the preceeding
methods,
the voltage
across
the
sample
cannot
be
arbitrarily
set.
In some cases,
as in measuring
capacitor
leakage,
these
methods
involve
much more time
than
if
a Larger
voltage
could
be applied.
In the
external
voltage
method
the Model
6LOB Fs used as a fast
picoammeter.
The unknown
resistance
sample
is
connected
to an external
known voltage
source
and the
current
through
the
sample
is measured,
Either
the normal
or fast
method
may be used.
The resistance
is
calculated
from
the readings.
NOTE
Discharge
any capacitor
before
removing
it
from
the
circuit.
Depressing
the ZERO CHECK Button
shorts
the
fnput
through
a
lo-megohm
resistor,
providing
a discharge
path.
b.
Normal
Method
(100
to 10"
ohm ranges).
1.
Follow
the
instructions
of paragraph
2-4.
Set the
controls
as follows:
METER Switch
Range Switch
To5 OHMS
Multiplier
Swftch
1
FEEDBACK Switch
NORMAL
ZERO CHECK Button
LOCK
Connect
the
resistance
sample
to the
INPUT Receptacle.
Unlock
the ZERO
CHECK Button.
Check zero
with
only
the ZERO CHECK Button,
Do not
open-circuit
the
Electrometer
on the OHMS ranges;
the
input
will
develop
a Large
voltage
due to its
constant
current
characteristic.
Keep the input
shorted
or the
ZERO CHECK
Button
locked.
2.
The full-scale
ohms range
is
the Range
Switch
setting
times
the
Multiplier
Switch
setting.
Use the
Largest
Multiplier
Switch
setting
possible
to obtain
the best
accuracy.
1064R
13